首页> 外文期刊>Bulletin of the Korean Chemical Society >Comprehension of the Response Time in a Microbial BOD Sensor(¥±)
【24h】

Comprehension of the Response Time in a Microbial BOD Sensor(¥±)

机译:微生物生化需氧量传感器响应时间的理解(¥±)

获取原文

摘要

A microbial BOD sensor using immobilized Hansenula anomala was prepared for the estimation of BOD. The sensor voltage was increased with time and increasing concentration of GGA when it was inserted in a sample solution. A linear relationship was obtained with a correlation coefficient, 0.998 between the concentration of standard GGA solutions and dV/dt by using the initial change of voltage, in which the response time was 20 min. It could be concluded that the oxidation of GGA conformed to a first-order kinetics. Therefore, the good linearities were also observed at various times. This sensor showed the best linearity at 30 min.
机译:制备了使用固定化汉逊酵母(Hansenula anomala)的微生物BOD传感器来估算BOD。当传感器电压插入样品溶液中时,传感器电压会随着时间增加并增加GGA的浓度。通过使用电压的初始变化(响应时间为20分钟),标准GGA溶液的浓度与dV / dt之间的线性关系为0.998,具有线性关系。可以得出结论,GGA的氧化符合一级动力学。因此,在不同时间也观察到良好的线性。该传感器在30分钟时显示出最佳的线性度。