首页> 外文期刊>Journal of Medical Physics/Association of Medical Physicists of India >Fast and accurate Monte Carlo modeling of a kilovoltage X-ray therapy unit using a photon-source approximation for treatment planning in complex media
【24h】

Fast and accurate Monte Carlo modeling of a kilovoltage X-ray therapy unit using a photon-source approximation for treatment planning in complex media

机译:使用光子源近似对复杂介质中的治疗计划进行快速准确的千伏X射线治疗仪蒙特卡洛建模

获取原文
       

摘要

To accurately recompute dose distributions in chest-wall radiotherapy with 120 kVp kilovoltage X-rays, an MCNP4C Monte Carlo model is presented using a fast method that obviates the need to fully model the tube components. To validate the model, half-value layer (HVL), percentage depth doses (PDDs) and beam profiles were measured. Dose measurements were performed for a more complex situation using thermoluminescence dosimeters (TLDs) placed within a Rando phantom. The measured and computed first and second HVLs were 3.8, 10.3 mm Al and 3.8, 10.6 mm Al, respectively. The differences between measured and calculated PDDs and beam profiles in water were within 2 mm/2% for all data points. In the Rando phantom, differences for majority of data points were within 2%. The proposed model offered an approximately 9500-fold reduced run time compared to the conventional full simulation. The acceptable agreement, based on international criteria, between the simulations and the measurements validates the accuracy of the model for its use in treatment planning and radiobiological modeling studies of superficial therapies including chest-wall irradiation using kilovoltage beam.Keywords: Fast Monte Carlo, kilovoltage therapy, modeling of external therapy, treatment planning
机译:为了用120 kVp千伏X射线精确地重新计算胸壁放射治疗中的剂量分布,提出了一种MCNP4C蒙特卡洛模型,该模型使用了一种快速方法,无需对管组件进行完全建模。为了验证模型,测量了半值层(HVL),深度剂量百分比(PDD)和光束轮廓。对于更复杂的情况,使用放置在Rando体模中的热发光剂量计(TLD)进行剂量测量。测量和计算的第一和第二HVL分别为3.8、10.3 mm Al和3.8、10.6 mm Al。对于所有数据点,在水中测量和计算的PDD和光束轮廓之间的差异在2 mm / 2%以内。在Rando模型中,大多数数据点的差异在2%以内。与常规的完全仿真相比,所提出的模型减少了大约9500倍的运行时间。根据国际标准,模拟和测量之间可接受的协议验证了该模型在表面计划(包括使用千伏束的胸壁照射的治疗计划)和放射生物学建模研究中使用的准确性。治疗,外部治疗建模,治疗计划

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号