首页> 外文期刊>Journal of Medical Physics/Association of Medical Physicists of India >Comparative study of convolution, superposition, and fast superposition algorithms in conventional radiotherapy, three-dimensional conformal radiotherapy, and intensity modulated radiotherapy techniques for various sites, done on CMS XIO planning system
【24h】

Comparative study of convolution, superposition, and fast superposition algorithms in conventional radiotherapy, three-dimensional conformal radiotherapy, and intensity modulated radiotherapy techniques for various sites, done on CMS XIO planning system

机译:在CMS XIO规划系统上进行的常规放疗,三维保形放疗和强度调制放疗技术在不同部位的卷积,叠加和快速叠加算法的比较研究

获取原文
           

摘要

The aim of this study is to compare the dosimetry results that are obtained by using Convolution, Superposition and Fast Superposition algorithms in Conventional Radiotherapy, Three-Dimensional Conformal Radiotherapy (3D-CRT), and Intensity Modulated Radiotherapy (IMRT) for different sites, and to study the suitability of algorithms with respect to site and technique. For each of the Conventional, 3D-CRT, and IMRT techniques, four different sites, namely, Lung, Esophagus, Prostate, and Hypopharynx were analyzed. Treatment plans were created using 6MV Photon beam quality using the CMS XiO (Computerized Medical System, St.Louis, MO) treatment planning system. The maximum percentage of variation recorded between algorithms was 3.7% in case of Ca.Lung, for the IMRT Technique. Statistical analysis was performed by comparing the mean relative difference, Conformity Index, and Homogeneity Index for target structures. The fast superposition algorithm showed excellent results for lung and esophagus cases for all techniques. For the prostate, the superposition algorithm showed better results in all techniques. In the conventional case of the hypopharynx, the convolution algorithm was good. In case of Ca. Lung, Ca Prostate, Ca Esophagus, and Ca Hypopharynx, OARs got more doses with the superposition algorithm; this progressively decreased for fast superposition and convolution algorithms, respectively. According to this study the dosimetric results using different algorithms led to significant variation and therefore care had to be taken while evaluating treatment plans. The choice of a dose calculation algorithm may in certain cases even influence clinical results.Keywords: Algorithm, conformity index, homogenity index, treatment planning system
机译:这项研究的目的是比较在不同部位使用常规放疗,三维保形放疗(3D-CRT)和强度调制放疗(IMRT)的卷积,叠加和快速叠加算法获得的剂量学结果,以及研究算法在站点和技术方面的适用性。对于每种常规,3D-CRT和IMRT技术,分析了四个不同的部位,分别是肺,食道,前列腺和下咽。使用CMS XiO(计算机医疗系统,密苏里州圣路易斯)的治疗计划系统,使用6MV光子束质量创建治疗计划。对于IMRT技术,在Ca.Lung情况下,算法之间记录的最大变异百分比为3.7%。通过比较目标结构的平均相对差,合格指数和均质指数进行统计分析。对于所有技术,快速叠加算法在肺和食道病例中均显示出极佳的效果。对于前列腺,叠加算法在所有技术中均显示出更好的结果。在下咽的常规情况下,卷积算法是好的。如果是Ca。肺,钙前列腺,食道钙和下咽钙,OARs通过叠加算法获得更多剂量;对于快速叠加算法和卷积算法,这分别逐渐减小。根据这项研究,使用不同算法的剂量学结果导致明显的差异,因此在评估治疗计划时必须格外小心。在某些情况下,剂量计算算法的选择甚至可能会影响临床结果。关键词:算法,一致性指标,均匀性指标,治疗计划系统

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号