首页> 外文期刊>Journal of Medical Physics/Association of Medical Physicists of India >Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators
【24h】

Evaluation methods for detecting changes in beam output and energy in radiation beams from high-energy linear accelerators

机译:用于检测来自高能线性加速器的辐射束的束输出和能量变化的评估方法

获取原文
       

摘要

There is need for simple methods for checking consistency of beam outputs and energy in linear accelerators used for radiotherapy. A method was designed by the department using perspex phantom with which the dosimetric data of two medical linear accelerators (Clinac 600 CD, Clinac 2300 CD) were evaluated over a period of 30 months. The efficacy of methods followed was checked. Routine beam consistency checks were designed for photon beams with 15 cm/ 5 cm depth ionizations in perspex phantom and variable depth combinations for electron beams. Calculated ionization ratios were compared with measured values to show their significance. The dose/MU for all radiation beams was maintained within 2% accuracy over the period of 30 months. Clinac 600 CD machine showed decreasing trend of cGy/MU, while Clinac 2300 CD showed increasing trend of cGy/MU over a period, which needed tuning of monitor chamber two times each. Tuning of output to achieve standard value was carried out once, for all electron energies when the output dose/MU exceeded 3%. During one week (June 2005), there were slight changes in electron energy detected using the ratio method, which did not recur anytime afterwards. The methods designed are adequate to find the consistency in the beam output and energies in the radiotherapy linacs.Keywords: Calibration of linacs, perspex phantoms, quality assurance methods
机译:需要用于检查用于放射疗法的线性加速器中的束输出和能量的一致性的简单方法。该部门设计了一种使用有机玻璃体模的方法,在30个月内评估了两种医用线性加速器(Clinac 600 CD,Clinac 2300 CD)的剂量学数据。检查了所采用方法的有效性。常规束一致性检查是针对在有机玻璃体模中具有15 cm / 5 cm深度电离的光子束和针对电子束的可变深度组合而设计的。将计算出的电离率与测量值进行比较,以显示其重要性。在30个月的时间内,所有辐射束的剂量/ MU均保持在2%的精度范围内。 Clinac 600 CD机在一段时间内显示cGy / MU下降的趋势,而Clinac 2300 CD机在一段时间内显示cGy / MU的上升趋势,这需要每次对监视室进行两次调谐。当输出剂量/ MU超过3%时,对所有电子能量进行一次输出调整以达到标准值。在一个星期内(2005年6月),使用比例法检测到的电子能量发生了微小变化,此后一直没有再出现。设计的方法足以在放射治疗直线加速器中找到光束输出和能量的一致性。关键词:直线加速器的校准,有机玻璃体模,质量保证方法

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号