Doppler Ultrasonographic, Oscillometric Sphygmomanometric, and Photoplethysmographic Techniques for Noninvasive Blood Pressure Measurement in Anesthetized Cats

机译:多普勒超声,示波血压计和光电容积描记技术用于麻醉猫的无创血压测量。

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Blood pressure (BP) measurements obtained using 3 indirect BP measuring instruments, a Doppler ultrasonic flowmeter, an oscillometric device, and a photoplethysmograph, were compared with direct arterial pressure measurements in 11 anesthetized cats. The standard deviation of the differences (SOD) between direct and indirect pressure measurements were not significantly different from each other ( P <.01), and ranged from 10.82 to 24.32 mm Hg. The high SDD values indicate that indirect BP estimates obtained with all these devices must be interpreted cautiously in individual patients. The mean errors (calculated as the sum of the differences between direct and indirect pressure measurements divided by the number of observations) of the 3 indirect devices were significantly different for systolic (SAP), diastolic (DAP), and mean (MAP) arterial pressures ( P <.05). The Doppler and photoplethysmographic devices had the highest overall accuracy, as indicated by mean error values of less than 10 mm Hg. Correlation coefficients varied from .88 to .96 for the Doppler flowmeter, and from .85 to .94 for the photoplethysmograph; for both devices, the regression line slopes were close to unity. The Doppler flowmeter detected a pulse under all experimental conditions. The photoplethysmograph was also efficient in obtaining BP measurements, obtaining over 90% of SAP, DAP, and MAP readings attempted. The oscillometric device was the least accurate, with mean error values varying from 10 to 22 mm Hg. Correlation coefficients were high (.90 to .94) for this device, but the slopes of the regression lines were 0.7 to 0.8, indicating a trend for increased error at higher BP. The oscillometric device tended to underestimate BP by increasing amounts as the BP increased. The oscillometric device was the least efficient device for obtaining BP measurements ( P <.01).
机译:将使用3台间接BP测量仪,多普勒超声流量计,示波仪和光电容积描记器获得的血压(BP)测量值与11只麻醉猫的直接动脉压测量值进行了比较。直接和间接压力测量之间的差异标准差(SOD)彼此之间无显着差异(P <.01),范围为10.82至24.32 mm Hg。高SDD值表明,必须谨慎解释所有患者使用所有这些设备获得的间接BP估计值。对于动脉收缩压(SAP),舒张压(DAP)和平均(MAP)动脉压,这3种间接装置的平均误差(用直接和间接压力测量值之差除以观察数得出的总和)显着不同。 (P <.05)。多普勒和光电容积描记仪具有最高的总体准确度,其平均误差值小于10 mm Hg。多普勒流量计的相关系数从0.88到0.96,光电容积描记器的相关系数从0.85到0.94。对于这两种设备,回归线斜率均接近于1。多普勒流量计在所有实验条件下均检测到脉冲。光电容积描记器在获得BP测量值方面也很有效,尝试获得90%以上的SAP,DAP和MAP读数。示波计精度最差,平均误差值在10到22 mm Hg之间。该设备的相关系数很高(.90至.94),但是回归线的斜率为0.7至0.8,这表明在较高BP时误差增加的趋势。示波设备往往会随着BP的增加而增加其数量,从而低估BP。示波测量仪是获得BP测量值的效率最低的设备(P <.01)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号