The Gore VBX stent graft (Gore Medical, Flagstaff, Ariz) provides a new option for branching of fenestrated aortic endografts. However, its modular stent structure has raised concerns about potential kinking at the interspace between stent rings if lateralizing force exists between the fenestration and target vessel orifice. We present a case of near-occlusion of a VBX celiac branch due to narrowing of this interspace identified at postoperative month 3. Although the Gore VBX offers several potential advantages as a branch endoprosthesis, its design poses the risk of unpredictable, delayed-onset kinking and raises concern for its use as a fenestrated endograft branch.
机译:戈尔VBX支架移植物(戈尔医疗公司,弗拉格斯塔夫,亚利桑那州)为分支开窗的主动脉内移植物提供了新的选择。然而,如果在开窗孔和目标血管孔之间存在侧向力,则其模块化的支架结构引起了人们对支架环之间间隙可能发生扭结的担忧。由于术后3个月发现的这个间隙变窄,我们提出了一个VBX腹腔分支几乎闭塞的病例。尽管Gore VBX作为分支内假体提供了一些潜在的优势,但其设计带来了无法预测的,延迟发作的扭曲的风险。并引起关注其用作有孔的内移植分支。