首页> 外文期刊>Journal of Radiation Research and Applied Sciences >A solution for 40K interference from K internal to the human body, in an underground high sensitivity whole body counter at the Waste Isolation Pilot Plant, Carlsbad, NM, USA
【24h】

A solution for 40K interference from K internal to the human body, in an underground high sensitivity whole body counter at the Waste Isolation Pilot Plant, Carlsbad, NM, USA

机译:在美国新墨西哥州卡尔斯巴德的废物隔离试验工厂的地下高灵敏度全身计数器中,解决了K干扰人体内部K的 40 K的解决方案

获取原文
           

摘要

The objective of the present work is to find a solution for the interference arising from 1461?keV photons emitted by 40 K in natural K internal to the human body in the detection of 241 Am and 210 Pb activities at the level of 37 mBq (1?pCi). The response of a broad energy Ge (BEGe) reference crystal to 1461?keV photons was simulated using GEANT4 code. Simulations were performed for a point source without and with shielding, as well as for a Bottle Manikin Absorption (BOMAB) phantom in the shielded whole-body counting chamber (WBC). The WBC facility is located underground at the Waste Isolation Pilot Plant, Carlsbad, NM, USA. The simulation results were discussed with regards to the interference in detecting the 241 Am and 210 ?Pb?at the level of 37 mBq (1?pCi). The specific conclusion is that the interference counts from 40 K activity at the natural levels present in the human body in typical clinical settings were found to be greater than the 210 Pb and 241 Am signal counts from 37 mBq (1?pCi), when a 38.1?mm diameter, 25.4?mm thick BEGe reference crystal was used. Our solution for minimizing the interference to the desired precision at the specified levels was found by sacrificing the broad energy response of the counting system. Highlights ? Detection of 37 mBq of 241 Am and 210 Pb and effect of 40 K interference. ? Simulated BEGe crystal response to 1461 keV photons of 40 K. ? Simulations done with point source, BOMAB phantom, with and without shielding. ? K interference counts exceeded the signal counts of 241 Am and 210 Pb. ? Counting time to 25 h or more is necessary to detect 37 mBq activity.
机译:本工作的目的是找到一种解决方案,以检测在37 mBq的水平下检测241 Am和210 Pb的活动时人体内部的天然K中40 K发射的1461?keV光子产生的干扰(1 ?pCi)。使用GEANT4代码模拟了宽能Ge(BEGe)参考晶体对1461?keV光子的响应。对不带屏蔽和带屏蔽的点源以及带屏蔽的全身计数室(WBC)中的瓶子人体模型吸收(BOMAB)体模进行了仿真。 WBC设施位于美国新墨西哥州卡尔斯巴德的废物隔离试验工厂的地下。讨论了仿真结果,涉及在37 mBq(1?pCi)的水平上检测241 Am和210?Pb?的干扰。具体结论是,在典型的临床环境中,在人体中自然水平下,来自40 K活性的干扰计数被发现大于来自37 mBq(1?pCi)的210 Pb和241 Am信号计数。使用直径38.1毫米,厚25.4毫米的BEGe参考晶体。通过牺牲计数系统广泛的能量响应,我们找到了在指定水平上将干扰最小化至所需精度的解决方案。强调 ?检测到241 Am和210 Pb的37 mBq和40 K干扰的影响。 ?模拟的BEGe晶体对40 K的1461 keV光子的响应。使用点光源,BOMB体模(带屏蔽和不带屏蔽)进行仿真。 ? K个干扰计数超过了241 Am和210 Pb的信号计数。 ?为了检测37 mBq的活性,需要计数到25小时或更长时间。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号