首页> 外文期刊>Journal of radiation research >Strength estimation of a moving 125Iodine source during implantation in brachytherapy: application to linked sources
【24h】

Strength estimation of a moving 125Iodine source during implantation in brachytherapy: application to linked sources

机译:近距离放射治疗中植入的移动式125碘源的强度估算:应用于链接源

获取原文
           

摘要

The speed of a source varies during brachytherapy implantations because it is moved manually. In the method proposed in this study, the strength of a source moving at an unknown and varying speed is estimated by performing short time measurements. The measurement should be completed while the source is in the region where the efficiency of the detector is constant. This was detailed in a previous work [9]. Additionally, the conditions for the counting time and the distance between the detector and the guiding needle were investigated in order to determine the source strength, regardless of the source speed. Four 125I seeds (STM1251, Bard Inc., Murray Hill, NJ, USA) were used in this study. One of them was used in the form of a ‘loose source' (i.e. single source). The other three sources were linked (with a distance between them of 10.5 mm) and separated by spacers (5.5 mm Standard Source Link, and 5.0 mm Extension Source Link, Bard Inc.). The strength of the loose source was 8.48 U. The linked sources had strengths of 8.38 U for Source #1, 8.43 U for Source #2, and 8.63 U for Source #3. The difference between the strengths was within 2%. Sources #1, #2 and #3 approached the detector in that order. The detector is described in the next subsection. The experimental set-up is shown in Fig. 1. The source was moved into the needle (#918201, Bard Inc.) by an applicator and a push rod (200-TPV, Mick Radio-Nuclear Instruments Inc., Mount Vernon, NY). The position and speed of the source was controlled with an electric actuator (EZS3D025-A, Oriental Motor Inc., Tokyo, Japan), which was connected to the push rod. In this experiment, the source was not implanted into a patient but only moved in the needle. The detector used was a plastic scintillator (G-tech Inc., Saitama, Japan). This was comprised of the scintillator (EJ200, Eljen Technology Inc., Sweetwater, TX), with dimensions of 80 mm × 50 mm × 20 mm, and the photomultiplier tube (H7416, Hamamatsu Photonics Inc., Shizuoka, Japan). At first, the dependence of the detector response on the source position along the needle was measured using the static sources. The static source measurement was performed for both the loose source and the linked sources (i.e. #1, #2 and #3). The measurements were also performed for both geometries, with and without the grid. Measurements of 1-s duration were conducted 10 times. Among the measured values in each run for the moving source, the maximum was utilized in further analyses, assuming that the source reached the position with the highest detector response when the maximum reading in each run was obtained. In order to investigate the accuracy of the result for the moving source, its ratio against the result for the static source (i.e. moving:static) was computed. This was based on the assumption that the correct result was obtained with the conventional static source measurement. A moving-to-static source ratio close to unity means that the measurement for the moving source was performed correctly. The dependence of the detector response on distance for the static loose source is shown in Fig. 2. The source position is an arbitrary value, which approximately corresponds to the distance from the end of the applicator exit to the source in Fig. 1a. The error bars indicate the standard deviation for 10 measurements. The counting rate of the background signal was 22 ± 2 cps, and this was subtracted from the measured response for the static source shown in Fig. 2. The reading for the detector-to-needle distance of 5 mm reached a maximum of 42 309 ± 724 cps at 46.5 mm. For the distance of 20 mm, the maximum was 16 314 ± 336 cps. Further measurements were performed for the distance of 5 mm to obtain higher counting rates with better statistics. An example of the raw data is shown in Fig. 4 for the source moving at 100 mm s–1 with the counting time of 20 ms in the geometry without the grid. The measurements for the moving sources were performed as five runs. The average and standard deviation of the maximum in each of the five runs are indicated in Table 1. The ratio against the result for the static source (i.e. moving:static) is shown in Fig. 5. The results of the moving source showed better agreement with those of the static source when the grid was not used. Assuming that the source speed in the implantation ranged up to 200 mm s–1, a counting time of 10 ms was applied. In this case, the counting rate in Fig. 3b was converted to the counting time of 10 ms for the static source. As shown in Fig. 5b, the moving-to-static ratio in 10 ms counting time at 200 mm s–1 was between 1.02 and 1.08 for the three sources.
机译:在近距离放射治疗植入期间,由于放射源的速度是手动移动的,因此其速度会有所不同。在这项研究中提出的方法中,通过执行短时测量来估算以未知且变化的速度移动的光源的强度。当源位于探测器效率恒定的区域时,应完成测量。先前的工作对此进行了详细介绍[9]。另外,研究了计数时间的条件以及探测器与导针之间的距离,以便确定源强度,而与源速度无关。在该研究中使用了四颗125 S种子(STM1251,Bard Inc.,Murray Hill,NJ,美国)。其中之一以“松散来源”(即单一来源)的形式使用。链接了其他三个信号源(它们之间的距离为10.5毫米),并用垫片(Bard Inc.的5.5毫米标准信号源链路和5.0毫米扩展信号源链路)隔开。松散源的强度为8.48U。链接源的强度对于源1是8.38 U,对于源2是8.43 U,对于源3是8.63U。强度之间的差异在2%以内。源#1,#2和#3以此顺序进入检测器。检测器将在下一部分中描述。实验装置如图1所示。通过施加器和推杆(200-TPV,Mick Radio-Nuclear Instruments Inc.,Mount Vernon,美国)将放射源移入针头(Brad Inc.,918918号),纽约)。源的位置和速度由电动执行器(EZS3D025-A,东方汽车公司,日本东京)控制,该执行器与推杆相连。在该实验中,放射源未植入患者体内,仅在针头中移动。所使用的检测器是塑料闪烁体(日本崎玉市G-tech Inc.)。它由尺寸为80 mm×50 mm×20 mm的闪烁体(EJ200,Eljen Technology Inc.,Sweetwater,TX)和光电倍增管(H7416,Hamamatsu Photonics Inc.,静冈,日本)组成。首先,使用静态源测量检测器响应对沿针的源位置的依赖性。对松散源和链接源(即#1,#2和#3)都执行了静态源测量。在有和没有网格的情况下,也对两种几何形状都进行了测量。 1-s持续时间的测量进行了10次。在移动源每次运行的测量值中,最大值用于进一步分析,假设在获得每次运行的最大读数时,源到达了具有最高检测器响应的位置。为了研究移动源的结果的准确性,计算了其与静态源的结果之比(即Moving:static)。这是基于以下假设:使用常规静态源测量可获得正确的结果。动静源比接近于1表示对动源的测量已正确执行。图2显示了检测器响应对静态松散源距离的依赖性。源位置是一个任意值,大约对应于图1a中从涂抹器出口端到源的距离。误差线表示10次测量的标准偏差。本底信号的计数速率为22±2 cps,这是从图2所示的静态源的测量响应中减去的。检测器到针距5 mm的读数最大为42 309。在46.5毫米处为±724厘泊。对于20 mm的距离,最大值为16314±336 cps。对5 mm的距离进行了进一步的测量,以获得更高的计数率和更好的统计数据。图4中显示了原始数据的一个示例,其中在没有网格的情况下,源以100 mm s –1 移动且计数时间为20 ms。对移动源的测量以五次运行进行。表1中列出了这五个运行中最大值的平均值和标准偏差。图5显示了静态源与静态源(即移动:静态)的结果之比。移动源的结果显示更好不使用网格时与静态源的协议一致。假设注入中的源速度范围高达200 mm s –1 ,则应用10 ms的计数时间。在这种情况下,将图3b中的计数率转换为静态源的10毫秒计数时间。如图5b所示,三个源在200 mm s -1 的10 ms计数时间内的动静比在1.02和1.08之间。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号