Evaluation of tensile strength and surface topography of orthodontic wires after infection control procedures: An in vitro study

机译:感染控制程序后对正畸钢丝的拉伸强度和表面形貌的评估:一项体外研究

获取原文
           

摘要

Aim:The aim of this study is to evaluate, the influence of four types of sterilization/disinfection procedures (autoclave, hot air oven, glutaraldehyde, and ultraviolet [UV] light) on the tensile strength and surface topography of three orthodontic wires (stainless steel (SS), titanium - molybdenum alloy [TMA], and cobalt chromium (CoCr)).Materials and Methods:Sample comprised of three types of 8 inches straight length segments of orthodontic wires. They were divided into three groups according to wire composition comprising of 50 samples each. Totally 50 samples of each group were then equally divided into five subgroups according to sterilization method. After sterilization and disinfection of the experimental group, surface topography was examined with scanning electron microscope (SEM) and tensile strength was tested using universal testing machine.Result:The results of this study show that the mean ultimate tensile strength (UTS) of SS wire after four sterilization procedures were similar to the control group (1845.815 ± 142.29 MPa). The mean UTS of TMA wire increases after four sterilization procedures when compared with the control group (874.107 ± 275.939 MPa). The mean UTS of CoCr wire remains same after UV light disinfection, but increases after other three sterilization procedures when compared with the control group (1449.759 ± 156.586 MPa). SEM photographs of the present study shows gross increase in pitting roughness of the surface topography of all the three types of wires after four types of sterilization.Conclusion:Orthodontists who want to offer maximum safety for their patients can sterilize orthodontic wires before placement, as it does not deteriorate the tensile strength and surface roughness of the alloys.
机译:目的:本研究的目的是评估四种消毒/消毒程序(高压灭菌器,热风烤箱,戊二醛和紫外线[UV]光)对三根正畸丝(不锈钢)的拉伸强度和表面形貌的影响钢(SS),钛钼合金[TMA]和钴铬(CoCr))。材料和方法:样品由三种类型的8英寸长的正畸线段组成。根据导线组成将它们分为三组,每组包括50个样品。然后根据灭菌方法将每组总共50个样品均等地分为五个亚组。实验组经过消毒灭菌后,用扫描电子显微镜(SEM)检查表面形貌,并用通用试验机测试拉伸强度。经过四次灭菌后的程序与对照组相似(1845.815±142.29 MPa)。与对照组相比,经过四个灭菌程序后,TMA焊丝的平均UTS值有所增加(874.107±275.939 MPa)。紫外线消毒后,CoCr焊丝的平均UTS保持不变,但与对照组相比(1449.759±156.586 MPa),在其他三种消毒程序后均增加。这项研究的SEM照片显示,四种类型的灭菌后,所有三种导线的表面形貌的点蚀粗糙度都有所增加。不会降低合金的拉伸强度和表面粗糙度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号