Comparison results between PTB and CIMM on the force transducer calibration under sinusoidal loading

机译:PTB和CIMM在正弦载荷下力传感器校准的比较结果

获取原文
           

摘要

Sinusoidal force is a typical kind of dynamic force applied widely in practice, and the sinusoidal force calibration systems are set up in Physikalisch-Technische Bundesanstalt(PTB) and Changcheng Institute of Metrology & Measurement(CIMM) respectively. To examine and compare the measurement methods for force transducer calibration, a comparison on sinusoidal force calibration is taken between PTB and CIMM. The experiment results show that the absolute value of the normalized error En between both sides is always less than 1, in the frequency range from 20Hz to 2000Hz, by using mass blocks A(nominal 1kg) and B(nominal 2kg). While the large deviation corresponding to the mass block C(nominal 4kg) at frequencies higher than 1000Hz may be contributed to some rocking effect at high frequency.
机译:正弦力是一种在实践中得到广泛应用的典型动态力,正弦力标定系统分别建立在德国物理技术联合会(PTB)和长城计量研究院(CIMM)上。为了检查和比较用于力传感器校准的测量方法,在PTB和CIMM之间进行了正弦力校准的比较。实验结果表明,使用质量块A(标称1kg)和B(标称2kg),在20Hz至2000Hz的频率范围内,两侧标准化误差En的绝对值始终小于1。虽然在高于1000Hz的频率下对应于质量块C(标称值为4kg)的较大偏差可能会导致高频下的某些摇摆效果。