Calibration of multi-channel pipettes using gravimetric method in accordance with the ISO 8655-6

机译:使用重量法根据ISO 8655-6校准多通道移液器

获取原文
获取外文期刊封面目录资料

摘要

A calibration system was set up for calibrating multi-channel pipettes using the gravimetric method in accordance with the ISO 8655-6. The calibration system consisted of a display unit, a motor control unit and a measuring unit. The measuring unit included a single weighing cell, twelve individual containers for receiving the dispensed volume from multi-channel pipettes and transport rails to automatically convey the containers to the weighing cell one by one. This paper described the measurement traceability for volume measurements to the SI unit of mass by using the calibration system and the validation of the calibration system for volume measurements. The calibration system was shown to be capable of measuring the dispensed volume from 30 μl to 1000 μl with the measurement uncertainty of 0.7 μl to 2.9 μl.
机译:根据ISO 8655-6,建立了用于使用重量分析法校准多通道移液器的校准系统。校准系统由显示单元,电机控制单元和测量单元组成。该测量单元包括一个称重传感器,十二个单独的容器(用于接收来自多通道移液器的分配体积)和运输导轨,以自动将容器一个一个地输送到称重传感器。本文描述了通过使用校准系统对体积测量到SI质量单位的测量可追溯性以及对体积测量校准系统的验证。校准系统显示能够测量30 µl至1000 µl的分配体积,测量不确定度为0.7 µl至2.9 µl。