首页> 外文期刊>Journal of nuclear medicine technology >Assessment of Camera Uniformity Using a Dynamic Line Phantom
【24h】

Assessment of Camera Uniformity Using a Dynamic Line Phantom

机译:使用动态线幻影评估相机均匀性

获取原文
           

摘要

A commercially available dynamic line phantom has been investigated for its suitability as a quality control tool for assessing scintillation camera uniformity. A nonuniformity was found to be introduced due to count rate losses. These losses were caused by the variation in the length of the line source exposed to a circular detector as the source was scanned across the camera face. Although the handling of the line source reduced the radiation hazard to which personnel would be exposed, the uniformity flood images produced by the phantom were unacceptable and could certainly not be used for uniformity correction in single-photon emission computed tomography (SPECT) imaging. The dynamic line phantom can also be programmed to produce several different line source resolution and contrast patterns. Under the test conditions some of these resolution patterns were too fine to be useful but could perhaps be of more value with more modern cameras.
机译:已经研究了市售的动态线体模是否适合用作评估闪烁相机均匀性的质量控制工具。发现由于计数率损失而引入了不均匀性。这些损失是由于在整个摄像机表面上扫描线源时,线源暴露于圆形检测器的长度变化而引起的。尽管对线源的处理减少了人员可能受到的辐射危害,但是幻影产生的均匀性泛光图像是不可接受的,并且肯定不能用于单光子发射计算机断层摄影(SPECT)成像中的均匀性校正。还可以对动态线体模进行编程,以产生几种不同的线源分辨率和对比度模式。在测试条件下,这些分辨率模式中的某些分辨率太好而无法使用,但对于更现代的相机可能更有价值。