Abnormal Cellular Repolarization due to Implantable Cardioverter Defibrillator Shock: A Case Report

机译:植入式心脏复律除颤器电击引起的异常细胞复极化:一例报告

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Implantable cardioverter-defibrillators (ICDs) have a considerable impact on the therapy of patients with threatening ventricular arrhythmias, significantly reducing the risk of sudden death. Even though newest devices have achieved a remarkable reliability, the delivery of inappropriate shocks is still possible. In this paper a singular case of T-wave oversensing is described, associated with a high number of inappropriate shocks and consequent rapid depletion of ICD battery. Persistent abnormalities of cellular repolarization following a first appropriate shock were believed to be at the origin of T-wave oversensing.J Med Cases. 2014;5(2):92-97doi: http://dx.doi.org/10.14740/jmc1642w
机译:植入式心脏复律除颤器(ICD)对具有威胁性的室性心律不齐的患者的治疗产生重大影响,从而大大降低了猝死的风险。即使最新的设备已经取得了显着的可靠性,但仍然可以传递不适当的冲击。在本文中,描述了奇异的T波过感应情况,伴有大量不适当的冲击和随之而来的ICD电池快速耗尽。第一次适当的电击后,持续的细胞复极化异常被认为是T波过度感应的根源。 2014; 5(2):92-97doi:http://dx.doi.org/10.14740/jmc1642w