“Simple mechanical devices did not improve pelvis positioning in AP pelvis radiographs for reliable assessment of the acetabular orientation”

机译:“简单的机械装置无法改善AP骨盆摄影中骨盆的位置,从而无法可靠地评估髋臼方位”

获取原文
           

摘要

Abstract BackgroundThe aim of this study was to develop two simple positioning devices for anteroposterior pelvis radiographs and to evaluate their effect on accuracy of the radiographs for assessment of the acetabular orientation compared with non-instrumented positioning.MethodsThe superior anterior iliac spines and the pubic symphysis were used as anatomical landmarks to obtain a horizontal orientation of the pelvis according to the anterior pelvic plane. Anteroposterior pelvis radiographs were taken of 11 human cadaveric pelvic bones with each of the positioning devices and without any device. Defined measurements were carried out to objectify the tilt and rotation of the pelvis and to assess the cross-over sign as well as the presence of the ischial spine sign. Computed tomography scans were performed as a standard of reference. Bland-Altman-Plots were used to compare the continuous measurement values and Cohen’s Kappa was applied for the categorical data. Intra- and inter-observer reliability was determined by the intraclass correlation coefficient and Cohen’s Kappa.ResultsThe mean values of the measurements showed a high variability. A low correlation of the measurement values was found between the radiographs of the different positioning methods and the computed tomography scans. The intra- and inter-observer reliability was good (Cohen’s Kappa 0.78 and 0.69; intraclass correlation coefficient 0.99 and 0.98).ConclusionThe use of positioning devices did not lead to more accurate anteroposterior pelvis radiographs compared to non-instrumented positioning. Simple positioning devices do not provide standardized anteroposterior pelvis radiographs for reliable assessment of the acetabular orientation.
机译:摘要背景本研究的目的是开发两种简单的骨盆前后位片定位装置,并评估其对骨盆位片准确性的影响,以评估髋臼定位与非器械位定位相比。方法The前上棘和耻骨联合用作解剖学界标以根据骨盆前平面获得骨盆的水平方向。使用每个定位设备和不使用任何设备的11具人体尸体骨盆骨拍摄前后骨盆X线照片。进行定义的测量以客观化骨盆的倾斜和旋转,并评估交叉征象以及坐骨脊柱征象的存在。以计算机断层扫描为标准。使用Bland-Altman-Plots比较连续测量值,并将Cohen的Kappa应用于分类数据。观察者内部和观察者之间的可靠性由类内相关系数和科恩氏Kappa决定。结果测量的平均值显示出较高的变异性。在不同定位方法的射线照片与计算机断层扫描之间发现测量值之间的相关性较低。观察者之间和观察者之间的可靠性都很好(Cohen Kappa值为0.78和0.69;组内相关系数为0.99和0.98)。结论与非器械式定位相比,使用定位装置不会导致更准确的前后骨盆X线片。简单的定位设备无法提供标准化的前后骨盆X线片,无法可靠地评估髋臼的方向。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号