首页> 外文期刊>Journal of Contemporary Brachytherapy >Inflatable multichannel rectal applicator for adaptive image-guided endoluminal high-dose-rate rectal brachytherapy: design, dosimetric characteristics, and first clinical experiences
【24h】

Inflatable multichannel rectal applicator for adaptive image-guided endoluminal high-dose-rate rectal brachytherapy: design, dosimetric characteristics, and first clinical experiences

机译:用于适应性影像引导的腔内高剂量率直肠近距离放射治疗的充气式多通道直肠涂药器:设计,剂量学特征和首次临床经验

获取原文
           

摘要

Purpose: To investigate the dosimetric results and first clinical experiences with a new designed balloon applicator with adjustable catheters for endoluminal brachytherapy for patients with locally advanced rectal cancer not undergoing surgery. Material and methods: The applicator consists of an inflatable rectal balloon with six attached Foley catheters used as guidance for the inserted brachytherapy plastic needles. The construction of the applicator and the dosimetric profile in terms of representative dose points in 0, 2, 5, 10 mm ipsilaterally and in 0 mm contralaterally are described. The first clinical outcomes in three patients are reported. Results: For all three patients, a reproducible dose gradient was achieved. The surface dose on the target side was 204 ± 19% of the normalized dose in 5 mm (100%) tissue depth, and 143 ± 8% in 2 mm and 64 ± 3% in 10 mm tissue depth, while the surface dose on the contra-lateral side was 20 ± 8%. After radiochemotherapy with 50 Gy external beam radiotherapy and concomitant administration of capecitabine, a HDR brachytherapy boost in 2-3 fractions of 7-10 Gy each was delivered. All patients achieved a clinical complete response 3 month after the treatment, and no major toxicity was observed. Conclusion: The use of the applicator was clinically feasible, and resulted in a stable and reproducible dose distribution. First clinical results are promising.
机译:目的:使用新型设计的带可调节导管的球囊涂药器进行剂量测定结果和首次临床经验,该球囊涂药器可用于未行手术的局部晚期直肠癌患者的腔内近距离治疗。材料和方法:涂药器由一个可充气的直肠气球组成,该气球具有六个连接的Foley导管,用作插入的近距离放射治疗塑料针的导向。描述了在同侧0、2、5、10 mm和对侧0 mm中代表剂量点的施药器结构和剂量分布图。报告了三例患者的首次临床结果。结果:对于所有三名患者,均达到了可重现的剂量梯度。在5 mm(100%)组织深度中,目标侧的表面剂量为标准剂量的204±19%,在2 mm组织中为143±8%,在10 mm组织深度中为64±3%。对侧为20±8%。在用50 Gy外部束放射疗法进行放射化学疗法并同时给予卡培他滨后,以2-3份7-10 Gy的2-3份进行HDR近距离放射治疗。所有患者在治疗后3个月均达到临床完全缓解,未观察到重大毒性反应。结论:使用涂药器在临床上是可行的,并导致稳定且可重复的剂量分布。最初的临床结果很有希望。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号