The influence of screw type, alloy and cylinder position on the marginal fit of implant frameworks before and after laser welding

机译:螺钉类型,合金和圆柱体位置对激光焊接前后植入物框架边缘配合的影响

获取原文
           

摘要

Misfit at the abutment-prosthetic cylinder interface can cause loss of preload, leading to loosening or fracture of gold and titanium screws. OBJECTIVES: To evaluate the influence of screw type, alloy, and cylinder position on marginal fit of implant frameworks before and after laser welding. METHODS: After Estheticone-like abutments were screwed to the implants, thirty plastic prosthetic cylinders were mounted and waxed-up to fifteen cylindrical bars. Each specimen had three interconnected prosthetic components. Five specimens were one-piece cast in titanium and five in cobalt-chromium alloy. On each specimen, tests were conducted with hexagonal titanium and slotted gold screws separately, performing a total of thirty tested screws. Measurements at the interfaces were performed using an optical microscope with 5mm accuracy. After sectioning, specimens were laser welded and new measurements were obtained. Data were submitted to a four-way ANOVA and Tukey's multiple comparisons test (a=0.05). RESULTS: Slotted and hexagonal screws did not present significant differences regarding to the fit of cylinders cast in titanium, either in one-piece casting framework or after laser welding. When slotted and hexagonal screws were tested on the cobalt-chromium specimens, statistically significant differences were found for the one-piece casting condition, with the slotted screws presenting better fit (24.13μm) than the hexagonal screws (27.93 μm). Besides, no statistically significant differences were found after laser welding. CONCLUSIONS: 1) The use of different metal alloys do exert influence on the marginal fit, 2) The slotted and hexagonal screws play the exclusive role of fixing the prosthesis, and did not improve the fit of cylinders, and 3) cylinder position did not affect marginal fit values.
机译:基台-假体圆柱体界面的配合不当会导致预紧力损失,从而导致金和钛螺钉松动或断裂。目的:评估螺钉类型,合金和圆柱体位置对激光焊接前后植入物框架边缘配合的影响。方法:将类似美甲酮的基台拧入植入物后,安装30个塑料假体圆柱体,并打蜡到15个圆柱棒上。每个标本具有三个相互连接的假体组件。五个试样是一件式铸造钛合金,另外五个是钴铬合金。在每个样品上,分别用六角形钛螺钉和带槽金螺钉进行测试,总共进行了30个测试螺钉。使用5mm精度的光学显微镜在界面上进行测量。切片后,将样品激光焊接并获得新的测量结果。将数据提交至四次方差分析和Tukey的多重比较检验(a = 0.05)。结果:无论是在整体铸造框架中还是在激光焊接后,一字形和六角形螺钉在钛铸件的圆柱体装配方面都没有显着差异。当在钴铬样品上测试开槽和六角螺钉时,发现一件式铸造条件在统计学上有显着差异,其中开槽螺钉比六角螺钉(27.93μm)具有更好的配合度(24.13μm)。此外,激光焊接后没有发现统计学上的显着差异。结论:1)使用不同的金属合金确实会对边缘配合产生影响,2)开槽和六角螺钉起固定假体的排他性作用,并没有改善圆柱体的配合,并且3)圆柱体位置没有改变影响边际拟合值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号