Evaluation of the Analytical Performance of the Olympus AU5421 Applied with Parameters Developed by the Toshiba 200FR for Measurement of Blood Ethanol Concentration

机译:使用Toshiba 200FR开发的参数对奥林巴斯AU5421的分析性能进行评估,该参数用于测量血液中的乙醇浓度

获取原文
           

摘要

Background An accurate determination of blood ethanol concentrations is important. To minimize ethanol degradation in blood samples, sodium fluoride (NaF) collection tubes have been recommended for use. In this study, we attempted to utilize the Olympus AU5421 chemistry analyzer for ethanol analysis based on the parameters established for the Toshiba 200FR. We also evaluated the effect of NaF collection tubes on ethanol concentrations. Methods The precision, linearity, accuracy, and carry-over rate of the AU5421 analyzer were evaluated. The results of analysis using the AU5421 and Abbott AxSYM analyzers were also compared. The effects of NaF collection tubes on ethanol concentrations in stored samples were measured. Results The AU5421 showed a good precision, linearity, accuracy, and carry-over rate. The ethanol concentrations were well correlated with the results obtained using the AxSYM. There was no statistically significant difference in blood ethanol concentrations between the samples collected in tubes with NaF and those collected in tubes without NaF. Conclusions Since the AU5421 showed excellent analytical performance, the AU5421 could be used as an alternative to AxSYM for the determination of blood ethanol concentrations. Our analysis also indicated that there is no need to use NaF collection tubes if blood ethanol concentrations are analyzed within 3 h after blood collection. We believe that the results obtained in this study will have important implications for the use of the AU5421 system to measure blood ethanol concentrations.
机译:背景技术准确测定血液中的乙醇浓度很重要。为了最大程度地减少血液样本中的乙醇降解,建议使用氟化钠(NaF)收集管。在这项研究中,我们尝试根据为Toshiba 200FR建立的参数,利用Olympus AU5421化学分析仪进行乙醇分析。我们还评估了NaF收集管对乙醇浓度的影响。方法对AU5421分析仪的精密度,线性,准确性和残留率进行评估。还比较了使用AU5421和Abbott AxSYM分析仪的分析结果。测量了NaF收集管对储存样品中乙醇浓度的影响。结果AU5421表现出良好的精度,线性,准确性和残留率。乙醇浓度与使用AxSYM获得的结果密切相关。在装有NaF的试管中收集的样品与不含NaF的试管中收集的样品之间的血液乙醇浓度没有统计学上的显着差异。结论由于AU5421具有出色的分析性能,因此AU5421可以替代AxSYM用于测定血液中的乙醇浓度。我们的分析还表明,如果在采血后3小时内分析血液乙醇浓度,则无需使用NaF采集管。我们相信,这项研究中获得的结果将对使用AU5421系统测量血液中乙醇浓度具有重要意义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号