首页> 外文期刊>Journal of clinical biochemistry and nutrition. >Performance evaluation of two immunoassays for 25-hydroxyvitamin D
【24h】

Performance evaluation of two immunoassays for 25-hydroxyvitamin D

机译:两种25-羟基维生素D免疫测定的性能评估

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Although immunoassays in measuring 25-hydroxyvitamin D [25(OH)D] have been improved recently, relatively large differences are still seen between results of 25(OH)D measured by immunoassays and by liquid chromatography-tandem mass spectrometry (LC-MS/MS). In the present studies, we compared two immunoassays with LC-MS/MS for measuring 25(OH)D concentrations. Concentrations of 25-hydroxyvitamin D2 [25(OH)D2] and 25-hydroxyvitamin D3 [25(OH)D3] in serum samples from 59 healthy subjects were measured by two immunoassays including Siemens ADVIA Centaur Vitamin D Total (Centaur) and Roche Elecsys Vitamin D Total (Elecsys) and LC-MS/MS. To determine the cross reactivity of Elecsys and Centaur toward 25(OH)D2, a dosage of 200,000?IU vitamin D2 was given after first sampling. Serum samples were obtained 30 days later and concentrations of 25(OH)D2 and 25(OH)D3 were measured again. The results showed poor agreement between the immunoassays and LC-MS/MS in 25(OH)D2 and 25(OH)D3 measurements. The percentage of 25(OH)D2 cross-reactivity was 45.3% for Centaur and 41.2% for Elecsys and there was no significant difference between Centaur and Elecsys. In conclusion, Centaur and Elecsys perform unsatisfactorily in measuring 25(OH)D levels, especially for 25(OH)D2 cross-reactivity. Therefore, clinicians need to be aware of the underestimation of vitamin D status when using these immunoassays for measuring individuals supplemented with vitamin D2.
机译:尽管最近已改进了用于测定25-羟基维生素D [25(OH)D]的免疫测定方法,但在通过免疫测定法和液相色谱-串联质谱法(LC-MS /多发性硬化症)。在本研究中,我们比较了两种采用LC-MS / MS进行25(OH)D浓度测定的免疫测定方法。 25-羟基维生素D 2 [25(OH)D 2 ]和25-羟基维生素D 3 [25(OH)D 的浓度> 59 的血清样本中的> 3 ]通过两种免疫测定法进行了测定,包括西门子ADVIA半人马维生素D总量(Centaur)和罗氏Elecsys维生素D总量(Elecsys)和LC-MS / MS。为了确定Elecsys和Centaur对25(OH)D 2 的交叉反应性,首次取样后应给予200,000?IU维生素D 2 。 30天后获得血清样品,并再次测量25(OH)D 2 和25(OH)D 3 的浓度。结果表明,在25(OH)D 2 和25(OH)D 3 测量中,免疫分析方法与LC-MS / MS之间的一致性差。 25(OH)D 2 的交叉反应百分率在Centaur中为45.3%,在Elecsys中为41.2%,Centaur和Elecsys之间没有显着差异。总之,Centaur和Elecsys在测量25(OH)D水平方面表现不尽人意,尤其是对于25(OH)D 2 交叉反应性。因此,临床医生在使用这些免疫测定法来测量补充了维生素D 2 的个体时,需要意识到维生素D状态的低估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号