Comparison of Fully Automated Urine Sediment Analyzers H800-FUS100 and Labumat-Urised with Manual Microscopy

机译:全自动显微镜下的全自动尿液沉积物分析仪H800-FUS100和Labumat的比较

获取原文
       

摘要

Recent technical developments have focused on the full automation of urinalyses, however the manual microscopic analysis of urine sediment is considered the reference method. The aim of this study was to compare the performances of the LabUMat-UriSed and the H800-FUS100 with manual microscopy, and with each other.MethodsThe urine sediments of 332 urine samples were examined by these two devices (LabUMat-UriSed, H800-FUS100) and manual microscopy.ResultsThe reproducibility of the analyzers, UriSed and Fus100 (4.1–28.5% and 4.7–21.2%, respectively), was better than that with manual microscopy (8.5–33.3%). The UriSed was more sensitive for leukocytes (82%), while the Fus-100 was more sensitive for erythrocyte cell counting (73%). There were moderate correlations between manual microscopy and the two devices, UriSed and Fus100, for erythrocyte (r = 0.496 and 0.498, respectively) and leukocyte (r = 0.597 and 0.599, respectively) cell counting however the correlation between the two devices was much better for erythrocyte (r = 0.643) and for leukocyte (r = 0.767) cell counting.ConclusionIt can be concluded that these two devices showed similar performances. They were time-saving and standardized techniques, especially for reducing preanalytical errors such as the study time, centrifugation, and specimen volume for sedimentary analysis; however, the automated systems are still inadequate for classifying the cells that are present in pathological urine specimens.
机译:最近的技术发展集中在尿液分析的完全自动化上,但是,手动显微镜分析尿沉渣被认为是参考方法。这项研究的目的是比较手动显微镜和LabUMat-UriSed和H800-FUS100的性能,方法是使用这两种设备(LabUMat-UriSed,H800-FUS100结果与手动显微镜相比,UriSed和Fus100分析仪的重现性(分别为4.1–28.5%和4.7–21.2%)要好于手动显微镜(8.5–33.3%)。 UriSed对白细胞更敏感(82%),而Fus-100对红细胞计数更敏感(73%)。手动显微镜检查与两个设备(UriSed和Fus100)之间的红细胞(分别为r = 0.496和0.498)和白细胞(分别为r = 0.597和0.599)细胞计数之间存在中等相关性,但是两个设备之间的相关性要好得多对于红细胞(r = 0.643)和白细胞(r = 0.767)细胞计数。结论可以得出结论,这两种设备表现出相似的性能。它们是节省时间和标准化的技术,尤其是用于减少分析前的误差,例如研究时间,离心分离和用于沉积分析的样品量;然而,自动化系统仍不足以对病理性尿液标本中存在的细胞进行分类。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号