Evaluation of fast spin echo MRI sequence for an MRI guided high intensity focused ultrasound system for in vivo rabbit liver ablation

机译:MRI引导的高强度聚焦超声系统在体内兔肝消融中快速自旋回波MRI序列的评估

获取原文

摘要

The effectiveness of magnetic resonance imaging (MRI) to monitor thermal lesions created by High Intensity Focused Ultrasound (HIFU) in rabbit liver in vivo is investigated. The MRI sequences of T1- weighted, and T2-weighted fast spin echo (FSE) were evaluated. The main goal in this paper was to find the range of repetition time (TR) and range of echo time (TE) which maximizes the contrast to noise ratio (CNR). An ultrasonic transducer operating at 2 MHz was used, which is navigated using a positioning device. With T1W FSE the range of TR under which CNR is maximized ranges from 400 to 900 ms. The maximum contrast measured is approximately 25. With T2W FSE the range of TE that establishes maximum contrast is between 40 ms and 80 ms, with CNR of approximately 14. T1W FSE is much better than T2W FSE in detecting thermal lesions in liver. Both T1W and T2 W FSE were proven successful to image thermal lesions created by HIFU in rabbit liver in vivo.
机译:研究了磁共振成像(MRI)监测高强度聚焦超声(HIFU)在兔肝体内产生的热损伤的有效性。评估了T1加权和T2加权快速自旋回波(FSE)的MRI序列。本文的主要目的是找到重复时间(TR)的范围和回声时间(TE)的范围,以使对比度与噪声比(CNR)最大化。使用工作在2 MHz的超声换能器,使用定位设备对其进行导航。使用T1W FSE时,CNR最大化的TR范围为400到900 ms。测得的最大对比度约为25。使用T2W FSE时,建立最大对比度的TE范围在40毫秒至80毫秒之间,CNR约为14。T1WFSE在检测肝脏的热损伤方面比T2W FSE好得多。事实证明,T1W和T2 W FSE都能成功地对HIFU在兔肝中体内产生的热损伤成像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号