In this study, we evaluated the performance of an Elekta linac in the delivery of gated radiotherapy. Delivery accuracy was examined with an emphasis on the impact of using short gating windows (low monitor unit beam-on segments) or long beam hold times. The performance was assessed using a 20cm by 20cm open field with the radiation delivered using a range of beam-on and beam-off time periods. Gated delivery measurements were also performed for two SBRT plans delivered using volumetric modulated arc therapy (VMAT). Tests included both free-breathing based gating (covering a variety of gating windows) and simulated breath-hold based gating. An IBA MatriXX 2D ion chamber array was used for data collection, and the gating accuracy at low MU was evaluated using gamma passing rates. For the 20 cm by 20 cm open field, the measurements generally showed close agreement between the gated and non-gated beam deliveries. Discrepancies, however, began to appear with a 5-to-1 ratio of the beam-off to beam-on times. The discrepancies observed for these tight gating windows can be attributed to the small number of monitor units delivered during each beam-on segment. Dose distribution analysis from the delivery of the two SBRT plans showed gamma passing rates (± 1%, 2%/1 mm) in the range of 95% to 100% for gating windows of 25%, 38%, 50%, 63%, 75%, and 83%. Using a simulated sinusoidal breathing signal with a 4 second period, the gamma passing rate of free-breathing gating and breath-hold gating deliveries were measured in the range of 95.7% to 100%. In conclusion, the results demonstrate that Elekta linacs can accurately deliver respiratory gated treatments for both free-breathing and breath-hold patients. Some caution should be exercised with the use of very tight gating windows.
机译:在这项研究中,我们评估了Elekta直线加速器在门控放射治疗中的性能。检查了传送精度,重点是使用较短的选通窗口(监视单元的低光束照射段)或较长的光束保持时间会产生影响。使用20cm x 20cm的开放视野评估性能,并使用一定范围的光束开启和关闭时间来传输辐射。还对使用容积调制弧光治疗(VMAT)交付的两个SBRT计划进行了门控交付测量。测试包括基于自由呼吸的门控(覆盖各种门控窗口)和基于模拟屏气的门控。使用IBA MatriXX 2D离子室阵列进行数据收集,并使用伽玛通过率评估低MU下的门控精度。对于20 cm x 20 cm的开放视野,测量通常显示出门控和非门控光束传输之间的一致性。但是,开始出现的偏离是光束关闭时间与光束打开时间的比例为5:1。这些紧密的选通窗口观察到的差异可以归因于在每个波束接通段期间交付的监视器单元数量少。从两个SBRT计划的交付中进行的剂量分布分析显示,对于25%,38%,50%,63%的门窗,伽玛通过率(±1%,2%/ 1 mm)在95%至100%的范围内,75%和83%。使用具有4秒钟周期的模拟正弦呼吸信号,测量的自由呼吸门控和屏气门控输送的伽马通过率在95.7%至100%的范围内。总之,结果表明,Elekta直线加速器可以为自由呼吸和屏气患者提供准确的呼吸门控治疗。使用非常紧密的门窗时应格外小心。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号