In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment

机译:EBT3 GAFCHROMIC膜在乳房IOERT治疗中的体内剂量测定和屏蔽盘对准验证

获取原文
           

摘要

Intraoperative electron radiation therapy (IOERT) cannot usually benefit, as conventional external radiotherapy, from software systems of treatment planning based on computed tomography and from common dose verify procedures. For this reason, in vivo film dosimetry (IVFD) proves to be an effective methodology to evaluate the actual radiation dose delivered to the target. A practical method for IVFD during breast IOERT was carried out to improve information on the dose actually delivered to the tumor target and on the alignment of the shielding disk with respect to the electron beam. Two EBT3 GAFCHROMIC films have been positioned on the two sides of the shielding disk in order to obtain the dose maps at the target and beyond the disk. Moreover the postprocessing analysis of the dose distribution measured on the films provides a quantitative estimate of the misalignment between the collimator and the disk. EBT3 radiochromic films have been demonstrated to be suitable dosimeters for IVD due to their linear dose-optical density response in a narrow range around the prescribed dose, as well as their capability to be fixed to the shielding disk without giving any distortion in the dose distribution. Off-line analysis of the radiochromic film allowed absolute dose measurements and this is indeed a very important verification of the correct exposure to the target organ, as well as an estimate of the dose to the healthy tissue underlying the shielding. These dose maps allow surgeons and radiation oncologists to take advantage of qualitative and quantitative feedback for setting more accurate treatment strategies and further optimized procedures. The proper alignment using elastic bands has improved the absolute dose accuracy and the collimator disk alignment by more than 50%.PACS number: 87.55.kh
机译:像常规的外部放射治疗一样,术中电子放射治疗(IOERT)通常不能受益于基于计算机断层扫描的治疗计划软件系统和常用的剂量验证程序。因此,体内膜剂量法(IVFD)被证明是评估传递给靶标的实际辐射剂量的有效方法。进行了一种在乳房IOERT期间进行IVFD的实用方法,以改善有关实际递送至肿瘤靶标的剂量以及屏蔽盘相对于电子束的对准的信息。已经在屏蔽盘的两侧放置了两个EBT3 GAFCHROMIC膜,以便获得目标处和盘外的剂量图。而且,对在薄膜上测量的剂量分布的后处理分析提供了对准直仪和圆盘之间未对准的定量估计。 EBT3放射致变色膜已被证明是适合IVD的剂量计,因为它们在规定剂量附近的窄范围内具有线性剂量-光学密度响应,并且具有固定到屏蔽盘的能力,而不会给剂量分布带来任何失真。脱色膜的离线分析允许绝对剂量的测量,这确实是对靶器官正确暴露以及对屏蔽下健康组织剂量估计的非常重要的验证。这些剂量图使外科医生和放射肿瘤学家能够利用定性和定量反馈来设置更准确的治疗策略和进一步优化的程序。使用弹性带正确对准可以将绝对剂量精度和准直器盘对准提高50%以上.PACS编号:87.55.kh

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号