首页> 外文期刊>Journal of applied clinical medical physics / >Shielding assessment of a mobile electron accelerator for intra‐operative radiotherapy
【24h】

Shielding assessment of a mobile electron accelerator for intra‐operative radiotherapy

机译:术中放射治疗用移动式电子加速器的屏蔽评估

获取原文
           

摘要

The purpose of this investigation is to measure, characterize and report stray photon leakage and scatter radiation measurements from the Mobetron, an intra‐operative electron linear accelerator designed for use in an operating room environment. The study is needed due to recent changes to the shielding design of the Mobetron, and also to provide specific information that may be required by regulation in various jurisdictions. An analysis is performed on a number of manufactured units to determine an average 3D stray photon radiation map. This information provides a basis for determining patient‐based load restrictions and/or additional shielding for the operating room. The data presented evaluates the number of procedures for which the Mobetron may be safely operated in a typical unshielded operating room. PACS number: 87.53.‐j; 87.56.‐v
机译:这项研究的目的是测量,表征和报告来自Mobetron的杂散光子泄漏和散射辐射测量结果,Mobetron是一种术中电子线性加速器,设计用于手术室环境。由于最近对Mobetron的屏蔽设计进行了更改,因此需要进行此项研究,并且还需要提供各个辖区的法规可能要求的特定信息。对多个制造单元进行分析,以确定平均3D杂散光子辐射图。该信息为确定基于患者的负荷限制和/或对手术室的附加屏蔽提供了基础。呈现的数据评估了Mobetron在典型的非屏蔽手术室中可以安全操作的程序数量。 PACS编号:87.53.-j; 87.56.-v