首页> 外文期刊>Journal of applied clinical medical physics / >Evaluation of PC‐ISO for customized, 3D printed, gynecologic 192Ir HDR brachytherapy applicators
【24h】

Evaluation of PC‐ISO for customized, 3D printed, gynecologic 192Ir HDR brachytherapy applicators

机译:针对定制的3D打印妇科192Ir HDR近距离放射治疗器的PC-ISO评估

获取原文
           

摘要

The purpose of this study was to evaluate the radiation attenuation properties of PC-ISO, a commercially available, biocompatible, sterilizable 3D printing material, and its suitability for customized, single-use gynecologic (GYN) brachytherapy applicators that have the potential for accurate guiding of seeds through linear and curved internal channels. A custom radiochromic film dosimetry apparatus was 3D-printed in PC-ISO with a single catheter channel and a slit to hold a film segment. The apparatus was designed specifically to test geometry pertinent for use of this material in a clinical setting. A brachytherapy dose plan was computed to deliver a cylindrical dose distribution to the film. The dose plan used an 192 Ir source and was normalized to 1500 cGy at 1 cm from the channel. The material was evaluated by comparing the film exposure to an identical test done in water. The Hounsfield unit (HU) distributions were computed from a CT scan of the apparatus and compared to the HU distribution of water and the HU distribution of a commercial GYN cylinder applicator. The dose depth curve of PC-ISO as measured by the radiochromic film was within 1% of water between 1 cm and 6 cm from the channel. The mean HU was ‐ 10 for PC-ISO and ‐ 1 for water. As expected, the honeycombed structure of the PC-ISO 3D printing process created a moderate spread of HU values, but the mean was comparable to water. PC-ISO is sufficiently water-equivalent to be compatible with our HDR brachytherapy planning system and clinical workflow and, therefore, it is suitable for creating custom GYN brachytherapy applicators. Our current clinical practice includes the use of custom GYN applicators made of commercially available PC-ISO when doing so can improve the patient's treatment.PACS number: none
机译:这项研究的目的是评估PC-ISO(一种可商购的,生物相容性,可消毒的3D打印材料)的辐射衰减特性,以及其对定制的一次性妇科(GYN)近距离放射治疗涂药器的适用性,这些涂药器具有进行精确指导的潜力通过线性和弯曲的内部通道传播种子。在PC-ISO中用一个导管通道和一个狭缝固定膜段,以3D打印定制的放射致变色膜剂量测定仪。该设备专门设计用于测试与该材料在临床环境中使用相关的几何形状。计算了近距离放射治疗剂量计划,以将圆柱状剂量分布传递给薄膜。剂量计划使用192 Ir辐射源,并在距通道1 cm处标准化为1500 cGy。通过将薄膜暴露量与在水中进行的相同测试进行比较来评估材料。从设备的CT扫描中计算出Hounsfield单位(HU)的分布,并将其与水的HU分布和商用GYN钢瓶施加器的HU分布进行比较。通过放射致变色膜测量的PC-ISO剂量深度曲线在距通道1 cm至6 cm之间的水的1%以内。 PC-ISO的平均HU为‐ 10,水的平均HU为-1。不出所料,PC-ISO 3D打印过程的蜂窝结构产生了适度的HU值分布,但平均值与水相当。 PC-ISO的水当量足以与我们的HDR近距离放射治疗计划系统和临床工作流程兼容,因此,它适合于创建定制的GYN近距离放射治疗应用器。我们目前的临床实践包括使用可商购的PC-ISO制成的定制GYN涂抹器,以改善患者的治疗情况。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号