Usefulness of direct‐conversion flat‐panel detector system as a quality assurance tool for high‐dose‐rate 192Ir source

机译:直接转换平板探测器系统作为高剂量率192Ir源的质量保证工具的实用性

获取原文
           

摘要

The routine quality assurance (QA) procedure for a high-dose-rate (HDR) 192 Ir radioactive source is an important task to provide appropriate brachytherapy. Traditionally, it has been difficult to obtain good quality images using the 192 Ir source due to irradiation from the high-energy gamma rays. However, a direct-conversion flat-panel detector (d-FPD) has made it possible to confirm the localization and configuration of the 192 Ir source. The purpose of the present study was to evaluate positional and temporal accuracy of the 192 Ir source using a d-FPD system, and the usefulness of d-FPD as a QA tool. As a weekly verification of source positional accuracy test, we obtained 192 Ir core imaging by single-shot radiography for three different positions (1300/1400/1500 mm) of a check ruler. To acquire images for measurement of the 192 Ir source movement distance with varying interval steps (2.5/5.0/10.0 mm) and temporal accuracy, we used the high-speed image acquisition technique and digital subtraction. For accuracy of the 192 Ir source dwell time, sequential images were obtained using various dwell times ranging from 0.5 to 30.0 sec, and the acquired number of image frames was assessed. Analysis of the data was performed using the measurement analysis function of the d-FPD system. Although there were slight weekly variations in source positional accuracy, the measured positional errors were less than 1.0 mm. For source temporal accuracy, the temporal errors were less than 1.0%, and the correlation between acquired frames and programmed time showed excellent linearity ( R 2 = 1 ). All 192 Ir core images were acquired clearly without image halation, and the data were obtained quantitatively. All data were successfully stored in the picture archiving and communication system (PACS) for time-series analysis. The d-FPD is considered useful as the QA tool for the 192 Ir source.PACS number: 87.56.Fc
机译:高剂量率(HDR)192 Ir放射源的常规质量保证(QA)程序是提供适当近距离放射治疗的重要任务。传统上,由于高能伽马射线的照射,很难使用192 Ir源获得高质量的图像。但是,直接转换平板检测器(d-FPD)使得确认192 Ir源的定位和配置成为可能。本研究的目的是使用d-FPD系统评估192 Ir源的位置和时间精度,以及d-FPD作为QA工具的实用性。作为源位置精度测试的每周验证,我们通过单次射线照相对检查尺的三个不同位置(1300/1400/1500毫米)获得了192 Ir核成像。为了获取具有可变间隔步长(2.5 / 5.0 / 10.0 mm)和时间精度的192 Ir源移动距离的测量图像,我们使用了高速图像采集技术和数字减法。为了获得192 Ir源的驻留时间的准确性,使用0.5到30.0 sec范围内的各种驻留时间获得了连续图像,并评估了获取的图像帧数。使用d-FPD系统的测量分析功能进行数据分析。尽管光源位置精度每周有微小变化,但测得的位置误差小于1.0 mm。对于源时间精度,时间误差小于1.0%,并且采集的帧与编程时间之间的相关性显示出极好的线性度(R 2 = 1)。清晰地获取了所有192个Ir核心图像,而没有图像光晕,并且定量获得了数据。所有数据已成功存储在图片存档和通信系统(PACS)中,以进行时间序列分析。 d-FPD被认为是对192 Ir源的QA工具有用.PACS编号:87.56.Fc

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号