Stability of the Helical TomoTherapy Hi·Art II detector for treatment beam irradiations

机译:螺旋TomoTherapy Hi·Art II探测器对治疗束照射的稳定性

获取原文
           

摘要

The Hi·Art II Helical TomoTherapy (HT) unit is equipped with a built-in onboard MVCT detector used for patient imaging and beam monitoring. Our aim was to study the detector stability for treatment beam measurements. We studied the MVCT detector response with the 6 MV photon beam over time, throughout short-term (during an irradiation) and long-term (two times 50 days) periods. Our results show a coefficient of variation ≤ 1 % for detector chambers inside the beam (excluding beam gradients) for short- and long-term response of the MVCT detector. Larger variations were observed in beam gradients and an influence of the X-ray target where degradation was found. The results assume that an ‘air scan’ procedure is performed daily to recalibrate the detector with the imaging beam. On short term, the detector response stability is comparable to other devices. Long-term measurements during two 50-day periods show a good reproducibility.PACS numbers: 87.55.ne, 87.55.Qr
机译:Hi·Art II螺旋层析疗法(HT)单元配备了内置的MVCT检测器,用于患者成像和射线束监视。我们的目的是研究治疗束测量的探测器稳定性。我们研究了短期(照射期间)和长期(两次50天)中6 MV光子束随时间的MVCT检测器响应。我们的结果表明,对于MVCT检测器的短期和长期响应,光束内部的检测器腔(不包括光束梯度)的变化系数≤1%。观察到光束梯度的较大变化以及发现劣化的X射线目标的影响。结果假设每天执行一次“空中扫描”程序,以用成像光束重新校准检测器。从短期来看,检测器的响应稳定性可与其他设备媲美。在两个50天的时间内进行长期测量显示出良好的重现性.PACS编号:87.55.ne,87.55.Qr