We report a case series of 8 toric intraocular lenses (IOLs) noted to have paracentral inherent opacities after implantation in uneventful phacoemulsification cataract surgery. Defective areas had a microbubble appearance embedded paracentrally in the lens material. All patients were followed up at 1?day, 1?week, and 3?months postoperatively; uncorrected distance visual acuity was unaffected in all cases. None of the 8 patients reported symptoms of increased glare or significant visual distortion postoperatively. It is likely that the defective areas were induced by the manufacturing process and less likely that they were the result of instrument manipulation or ophthalmic viscosurgical device use. To our knowledge, this is the first report of defective toric IOLs with no effects on visual acuity.
机译:我们报告一例8复曲面人工晶状体(IOLs)的病例系列,这些病例在无畸形超声乳化白内障手术中植入后具有中心旁固有的混浊。缺陷区域在透镜材料的中央附近嵌有微泡外观。所有患者均在术后1天,1周和3个月进行随访。未经矫正的远距离视力在所有情况下均不受影响。 8例患者中无一例术后出现眩光增加或明显视力扭曲的症状。缺陷区域很可能是由制造过程引起的,而不太可能是仪器操作或眼科内窥镜手术设备使用的结果。据我们所知,这是复曲面人工晶状体缺陷的第一份报告,对视敏度没有影响。