An assessment of the performance of the Baxter elastomeric (LV10) Infusor? pump under hyperbaric conditions

机译:百特弹性体(LV10)输液器的性能评估?高压条件下的泵

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Introduction There are limited data on the use of elastomeric infusion pumps during hyperbaric oxygen treatment. Aim This study evaluated the flow rate of the Baxter elastomeric LV10 Infusor? pump under normobaric (101.3 kPa and three hyperbaric conditions of 203 kPa, 243 kPa and 284 kPa. Methods Elastomeric pumps were secured to participants in the same manner as for a typical patient, except that a container collected the delivered antibiotic solution. Pumps and tubing were weighed before and after the test period to determine volume delivered and to calculate flow rates at sea level and the three commonly used hyperbaric treatment pressures at two different time periods, 0–2 hours (h) and 19–21 h into the infusion. Results The mean flow rates in ml·h?1 (SD) were: 9.5 (0.4), 10.3 (0.6), 10.4 (0.6), 10.4 (0.5) at 0–2 h and 10.5 (1.0), 12.2 (0.6), 9.4 (0.5), 10.3 (0.9) at 19–21 h for the normobaric, 203 kPa, 243 kPa and 284 kPa conditions respectively. There was no significant association between flow rate and time period ( P = 0.166) but the 203 kPa flow rates were significantly faster than the other flow rates ( P = 0.008). In retrospect, the 203 kPa experiments had all been conducted with the same antibiotic solution (ceftazidime 6 g). Repeating that experimental arm using flucloxacillin 8 g produced flow rates of 10.4 (0.8) ml·h?1, with no significant associations between flow rate and time period ( P = 0.652) or pressure ( P = 0.705). Conclusion In this study, the flow rate of the Baxter LV10 Infusor? device was not significantly affected by increases in ambient pressure across the pressure range of 101.3 kPa to 284 kPa, and flow rates were generally within a clinically acceptable range of 9–12 ml·hr?1. However, there was evidence that the specific antibiotic solution might affect flow rates and this requires further study.
机译:简介在高压氧治疗期间使用弹性体输液泵的数据有限。目的本研究评估了百特弹性体LV10输液器的流速。在常压(101.3 kPa和203 kPa,243 kPa和284 kPa的三种高压条件下)泵。方法除了使用容器收集递送的抗生素溶液外,以与典型患者相同的方式将弹性体泵固定到参与者身上。在测试期间之前和之后分别称重,以确定输液量并计算海平面的流速以及输液的两个不同时间段(输注0–2小时(h)和19–21 h)的三种常用高压治疗压力。结果以ml·h?1(SD)为单位的平均流速为:0–2 h时为9.5(0.4),10.3(0.6),10.4(0.6),10.4(0.5)和10.5(1.0),12.2(0.6) ,在常压,203 kPa,243 kPa和284 kPa的常压下,在19–21 h分别为9.4(0.5),10.3(0.9),流速和时间段之间无显着关联(P = 0.166),但203 kPa流速显着高于其他流速(P = 0.008)。回顾一下,203 kPa实验中所有用相同的抗生素溶液(头孢他啶6 g)进行。重复使用氟氯西林8 g进行实验,产生的流速为10.4(0.8)ml·h?1,流速与时间段(P = 0.652)或压力(P = 0.705)之间没有显着关联。结论在本研究中,Baxter LV10输液器的流量?在101.3 kPa至284 kPa的压力范围内,环境压力的增加不会明显影响该设备,并且流速一般在9-12 ml·hr?1的临床可接受范围内。但是,有证据表明特定的抗生素溶液可能会影响流速,这需要进一步研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号