Preconcentration of heavy metals on oxides of cerium and zirconium and their determination by FAAS

机译:铈和锆的氧化物上重金属的预富集及其FAAS测定

获取原文
           

摘要

The use of commercial oxides of cerium and zirconium as potential sorbents in a column system was investigated. Efficacy of preconcentration of Cd, Co, Cr, Cu, Fe, Mn, Ni and Pb ions on CeO2 and ZrO2, prior to their determination by flame atomic absorption spectrometry (FAAS) was described. The influences of experimental parameters including pH of the sample solution and sample volume from pure multielement aqueous solutions of known concentration of analytes on the preconcentration efficiency have been investigated, applying the column method. Under optimum conditions (pH; 8, preconcentration factor; 10) the recovery values for CeO2 were: 87 % (Cd), 96 % (Co), 89 % (Cr), 104 % (Cu), 102 % (Fe), 74 % (Mn), 102 % (Ni) and 99 % (Pb); while using ZrO2 the recoveries were: 89 % (Cd), 97 % (Co), 86 % (Cr), 103 % (Cu), 105 % (Fe), 87 % (Mn), 103 % (Ni) and 95 % (Pb). Limits of detection were between 3.2-26.3 μg/L by using CeO2 as sorbent, as well, between 2.5-24.1 μg/L by using ZrO2 as sorbent. The precision expressed as relative standard deviation (R.S.D.) obtained for 0.1 mg/L of Cd, Co, Cr, Cu, Fe, Mn, Ni and Pb were 7 % for CeO2 and 5 % for ZrO2, calculated from eight measurements. Moreover, the sorption capacity of the CeO2 and ZrO2 were compared with other sorbents, and it was found that both sorbent offer good sorption capacity values and could be applicable to more analytes.
机译:研究了铈和锆的市售氧化物在色谱柱系统中作为潜在吸附剂的用途。描述了在通过火焰原子吸收光谱法(FAAS)测定之前,将Cd,Co,Cr,Cu,Fe,Mn,Ni和Pb离子预浓缩在CeO2和ZrO2上的功效。应用柱法研究了实验参数,包括样品溶液的pH值和已知分析物浓度的纯多元素水溶液的样品体积对预浓缩效率的影响。在最佳条件下(pH; 8,预浓缩因子; 10),CeO2的回收值为:87%(Cd),96%(Co),89%(Cr),104%(Cu),102%(Fe), 74%(Mn),102%(Ni)和99%(Pb);使用ZrO2时的回收率为:89%(Cd),97%(Co),86%(Cr),103%(Cu),105%(Fe),87%(Mn),103%(Ni)和95 %(Pb)。通过使用CeO2作为吸附剂,检测限在3.2-26.3μg/ L之间,通过使用ZrO2作为吸附剂,检测限在2.5-24.1μg/ L之间。由八次测量计算得出的0.1 mg / L的Cd,Co,Cr,Cu,Fe,Mn,Ni和Pb的相对标准偏差(R.S.D.)的精度对于CeO2为<7%,对于ZrO2为<5%。此外,将CeO2和ZrO2的吸附容量与其他吸附剂进行了比较,发现这两种吸附剂均具有良好的吸附容量值,可用于更多的分析物。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号