Comparison of peri-implant bone formation around injection-molded and machined surface zirconia implants in rabbit tibiae

机译:兔胫骨注射成型和机械加工表面氧化锆植入物周围植入物周围骨形成的比较

获取原文
   

获取外文期刊封面封底 >>

       

摘要

The aim of this study was to compare osseointegration and surface characteristics of zirconia implants made by the powder injection molding (PIM) technique against those made by the conventional milling procedure in rabbit tibiae. Surface characteristics of 2 types of implants were evaluated. Sixteen rabbits received 2 types of external hex implants with similar geometry, either machined zirconia implants or PIM zirconia implants, in the tibiae. Removal torque tests and histomorphometric analyses were performed. The roughness of the PIM zirconia implants was higher than that of machined zirconia implants. The PIM zirconia implants exhibited significantly higher bone-implant contact and removal torque values than the machined zirconia implants (p <0.001). The osseointegration of the PIM zirconia implant is promising, and PIM, using the roughened mold etching technique, can produce substantially rougher surfaces on zirconia implants.
机译:这项研究的目的是比较通过粉末注射成型(PIM)技术制成的氧化锆植入物与通过常规铣削方法在兔胫骨中制成的氧化锆植入物的骨整合和表面特性。评价了两种类型的植入物的表面特性。 16只兔子在胫骨中接受了两种类型的具有相似几何形状的外部六角形植入物,无论是机加工氧化锆植入物还是PIM氧化锆植入物。进行了去除扭矩测试和组织形态分析。 PIM氧化锆植入物的粗糙度高于机加工氧化锆植入物的粗糙度。 PIM氧化锆植入物显示出比机加工氧化锆植入物高得多的骨-植入物接触和去除扭矩值(p <0.001)。 PIM氧化锆植入物的骨整合是有前途的,并且使用粗糙化的模具蚀刻技术的PIM可以在氧化锆植入物上产生实质上更粗糙的表面。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号