首页> 外文期刊>The Turkish journal of pediatrics. >Diagnosis of cystic fibrosis with chloride meter (Sherwood M926S chloride analyzer?) and sweat test analysis system (CFΔ collection system?) compared to the Gibson Cooke method
【24h】

Diagnosis of cystic fibrosis with chloride meter (Sherwood M926S chloride analyzer?) and sweat test analysis system (CFΔ collection system?) compared to the Gibson Cooke method

机译:与吉布森·库克方法相比,使用氯化物测量仪(Sherwood M926S氯化物分析仪?)和汗液测试分析系统(CFΔ采集系统?)诊断囊性纤维化

获取原文
获取外文期刊封面目录资料

摘要

Our aim was to compare sweat chloride values obtained by GC method with other sweat test methods in patients diagnosed with CF and whose CF diagnosis had been ruled out. We wanted to determine if the other sweat test methods could reliably identify patients with CF and differentiate them from healthy subjects. Chloride concentration was measured with GC method, chloride meter and sweat test analysis system; also conductivity was determined with sweat test analysis system. Forty eight patients with CF and 82 patients without CF underwent the sweat test, showing median sweat chloride values 98.9 mEq/L with GC method, 101 mmol/L with chloride meter, 87.8 mmol/L with sweat test analysis system. In non-CF group, median sweat chloride values were 16.8 mEq/L with GC method, 10.5 mmol/L with chloride meter, and 15.6 mmol/L with sweat test analysis system. Median conductivity value was 107.3 mmol/L in CF group and 32.1 mmol/L in non CF group. There was a strong positive correlation between GC method and the other sweat test methods with a statistical significance (r=0.85) in all subjects. Sweat chloride concentration and conductivity by other sweat test methods highly correlate with the GC method. We think that the other sweat test equipments can be used as reliably as the classic GC method to diagnose or exclude CF.
机译:我们的目的是比较GC方法和其他汗液测试方法在诊断为CF且已被CF诊断的患者中获得的汗液氯化物值。我们想确定其他汗液测试方法是否可以可靠地识别CF患者并将他们与健康受试者区分开。用气相色谱法,氯化物测定仪和汗液测试分析系统测量氯离子浓度;用电导率测试分析系统测定电导率。对48位CF患者和82位无CF患者进行了汗液测试,GC法显示汗液中位数为98.9 mEq / L,氯计显示101 mmol / L,汗液分析系统中为87.8 mmol / L。在非CF组中,汗液氯化物的中值GC方法为16.8 mEq / L,氯离子计为10.5 mmol / L,汗液测试分析系统为15.6 mmol / L。 CF组电导率中位数为107.3 mmol / L,非CF组电导率中位数为32.1 mmol / L。在所有受试者中,GC方法与其他汗液测试方法之间均具有很强的正相关性,具有统计学意义(r = 0.85)。其他汗液测试方法的汗液氯化物浓度和电导率与GC方法高度相关。我们认为其他汗液测试设备可以像传统的GC方法一样可靠地用于诊断或排除CF。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号