首页> 外文期刊>Urology Case Reports >Failure of the inflatable penile prosthesis due to abnormal folding of a low-profile reservoir – A selected case from an overall series and systematic review
【24h】

Failure of the inflatable penile prosthesis due to abnormal folding of a low-profile reservoir – A selected case from an overall series and systematic review

机译:低矮型水箱的异常折叠导致可充气的阴茎假体失败-从整个系列和系统评价中选择的一个案例

获取原文
   

获取外文期刊封面封底 >>

       

摘要

We present a case from a running series of inflatable penile prosthesis failure due to improper folding of the Conceal? reservoir. The Conceal? Low-Profile reservoir gained popularity due to claims of improved cosmesis and ease of implantation. As the number of patients receiving this and other low-profile reservoirs increases, it is imperative to review and document any novel complications. While the Conceal? reservoir may be preferred in ectopic placement, it may be more prone to fluid lockout facilitated by conformational change. Our review did not identify prior reports of improper folding, which we believe is unique to these low-profile reservoirs.
机译:我们介绍了由于隐匿性折叠不当而导致的一系列可充气的阴茎假体失败的案例。水库。隐藏?低断面水库因具有改善美容效果和易于植入的优点而广受欢迎。随着接受这种和其他低调储液器的患者数量的增加,必须检查并记录任何新的并发症。而隐秘?在异位放置中可能首选储液罐,它可能更容易因构象变化而导致液体闭锁。我们的审查未发现先前有关不当折叠的报道,我们认为这些是这些低调储层所独有的。