The effect of changes to MOLLI scheme on T1 mapping and extra cellular volume calculation in healthy volunteers with 3 tesla cardiovascular magnetic resonance imaging

机译:3 tesla心血管磁共振成像对健康志愿者的MOLLI方案更改对T1定位和额外细胞体积计算的影响

获取原文
           

摘要

>>Background: Diffuse myocardial fibrosis may be quantified with magnetic resonance (MR) by calculating extracellular volume (ECV) fraction from native and post-contrast T1 values. The ideal modified look-locker inversion recovery (MOLLI) sequence for deriving T1 values has not been determined. This study aims to establish if systematic differences exist between suggested MOLLI schemes.>>Methods: Twelve phantom gels were studied with inversion recovery spin echo MR at 3.0 tesla to determine reference T1. Gels were then scanned with six MOLLI sequences (3s)3b(3s)5b; 4b(3s)3b(3s)2b; 5b(3s)3b with flip angles of both 35° and 50° at a range of heart rates (HRs). In 10 healthy volunteers MOLLI studies were performed on two separate occasions. Mid ventricular native and post contrast T1 was measured and ECV (%) calculated.>>Results: In phantoms, the co-efficient of variability at simulated HR [40-100] with a flip angle of 35° ranged from 6.77 to 9.55, and at 50° from 7.71 to 11.10. T1 was under-estimated by all MOLLI acquisitions. Error was greatest with longer T1, and increased as HR increased. The 10 volunteers had normal MR studies. Native T1 time was similar for all acquisitions but highest with the 5b(3s)3b 35° scheme (1,189.1±33.46 ms). Interstudy reproducibility was similar for all MOLLIs.>>Conclusions: The 5b(3s)3b MOLLI scheme agreed best with reference T1, without statistical difference between the six schemes. The shorter breath-hold time of 5b(3s)3b scheme may be preferable in clinical studies and warrants further investigation.
机译:> >背景:弥散性心肌纤维化可通过磁共振(MR)通过从天然T1值和对比后T1值计算细胞外体积(ECV)分数来量化。尚未确定用于导出T1值的理想修改后的外观锁反转恢复(MOLLI)序列。这项研究旨在确定建议的MOLLI方案之间是否存在系统差异。 > >方法:研究了十二个幻像凝胶,并在3.0 tesla的反向恢复自旋回波MR中确定了参考值T1。然后用六个MOLLI序列(3s)3b(3s)5b扫描凝胶; 4b(3s)3b(3s)2b; 5b(3s)3b,在一定心率(HRs)范围内的翻转角均为35°和50°。在10个健康志愿者中,MOLLI研究分别进行了两次。 > >结果:在幻影中,模拟心率[40-100]的变异系数与35°的翻转角在6.77至9.55之间,而50°的翻转角在7.71至11.10之间。 T1被所有MOLLI收购低估了。随着T1的延长,误差最大,并且随着HR的增加而增加。这10名志愿者的MR研究正常。所有采集的原始T1时间均相似,但最高5b(3s)3b 35°方案(1,189.1±33.46 ms)。结论: 5b(3s)3b MOLLI方案与参考文献T1最佳吻合,而这六种方案之间在统计学上无差异。 > >结论: 5b(3s)3b方案的屏气时间较短可能是临床研究中最好的方法,值得进一步研究。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号