Context:Point-of-care testing (POCT) for blood gas and chemistries is widely adopted in the emergency department (ED) for fast turnaround and critical medical decisions.Objective:To assess performance of ED physicians compared to laboratory technologists carrying out i-STAT analysis. Impact of user skill on POCT performance was also studied.Design:Over a 3-month period, ED physicians performed tests with their i-STAT unit with parallel blood specimens (n=60) sent to the central laboratory to be tested on another i-STAT unit and laboratory analysers (blood gas and chemistries). Comparisons to laboratory results (obtained with ABL 520 and Beckman-Coulter LX20PRO) were carried out. Two operators from ED and central laboratory performed side by side, repeated tests on the i-STAT units. Their overall within-run imprecision were compared. The analytical performance of i-STAT was further assessed by another laboratory technologist to verify its capability to deliver good results.Results:Platform comparisons show...
机译:背景:急诊室(ED)广泛采用了针对血气和化学物质的即时检验(POCT),以快速完成周转和关键医疗决策。目的:与进行i-操作的实验室技术人员相比,评估ED医师的表现STAT分析。设计:在3个月的时间里,ED医师使用i-STAT单元进行了测试,将i-STAT平行血液样本(n = 60)发送到中央实验室,在另一台i -STAT单元和实验室分析仪(气体和化学药品)。与实验室结果(用ABL 520和Beckman-Coulter LX20PRO获得)进行了比较。 ED和中央实验室的两名操作员对i-STAT单元并排进行了重复测试。比较了他们的整体运行中不精确度。另一位实验室技术人员进一步评估了i-STAT的分析性能,以验证其提供良好结果的能力。结果:平台比较显示...

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号