Validation of whole room indirect calorimeters: refinement of current methodologies

机译:整个房间间接量热计的验证:当前方法的完善

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Whole room indirect calorimeter (WRIC) validation techniques consist of propane combustion (PC) or infusion of mixed carbon dioxide (CO 2 ) and nitrogen (N 2 ) by a precision blender (PB). To determine the best method, PC of 6, 10, 22?¢????h and PB infusions of 6, 10, and 14?¢????h, were conducted. The 14?¢????h infusion consisted of two metabolic settings. Energy expenditure (EE; kJ), ventilation (V; liters/min) of oxygen (VO 2 ), VCO 2 , and respiratory quotient (VCO 2 /VO 2 ) obtained from the WRIC were extrapolated to the respective test durations and compared to similarly calculated values. Moreover, accurate equations (AE) were derived to correct infusions for additional N 2 . As a final evaluation of a PC validated WRIC, weight maintenance (WM), energy balance (EB), respiratory quotient (RQ), and food quotients (FQ) were determined in 22 subjects who had repeat 24?¢????h EE measurements. Statistical analyses ( P ????
机译:整个房间的间接热量计(WRIC)验证技术包括丙烷燃烧(PC)或通过精密混合器(PB)注入混合的二氧化碳(CO 2)和氮(N 2)。为了确定最佳方法,进行了6、10、22°h的PC灌注和6、10和14°h的PB灌注。 14小时的输注包括两个代谢环境。从WRIC获得的能量消耗(EE; kJ),通气量(V;升/分钟)(VO 2),VCO 2和呼吸商(VCO 2 / VO 2)被外推到相应的测试时间并与计算得出的值类似。此外,导出了精确的方程式(AE)来纠正其他N 2的输液。作为PC验证的WRIC的最终评估,确定了22名重复24 h≥h的受试者的体重维持(WM),能量平衡(EB),呼吸商(RQ)和食物商(FQ)。 EE测量。进行统计分析(P≤0.05)(SPSS,版本23)。 6 h后,PC和化学计量之间存在RQ的显着差异。其余PC测试的误差范围为1.5%±1.5%±2.4%(VCO 2)至2.8%±4.6% (EE)。当与WRIC进行比较时,所有6到10 h PB输注的未经校正的代谢参数均与?12.8 errors12.8 errors±??? 1.6 1.6有显着差异。 VO 2)为6.0%±2.8%(RQ)。 AE将误差的幅度降低到¢±12.4(±1.2)±1.5(RQ)至2.2 2.2±±3.0%(RQ) 。具有两种设置的PB输液显示出相似的性能。在受试者中,WM,EB,RQ或FQ没有差异。总之,10 h的PC测试足以验证WRIC。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号