首页> 外文期刊>Patient Safety in Surgery >Mechanical failure of a femoral lengthening nail: a case report
【24h】

Mechanical failure of a femoral lengthening nail: a case report

机译:股骨加长指甲的机械故障:一例报告

获取原文
   

获取外文期刊封面封底 >>

       

摘要

This case report describes the mechanical failure of a femoral lengthening nail in order to allow early recognition and prevent its occurrence. A femoral lengthening nail was used to achieve goal distraction length over eight weeks and sustained a mechanical failure during progressive weightbearing. Re-distraction was attempted with subsequent non-weight-bearing status. Although the patient was not able to attain the original goal length, radiographs demonstrated stable shortening of the patient’s right lower extremity along with early signs of incomplete regenerate bone. She eventually received exchange nailing with bone graft. Post-operative protocol and weight-bearing status should be further researched and standardized after re-evaluating the mechanical stiffness of the intramedullary nail construct.
机译:该病例报告描述了股骨加长指甲的机械故障,以便尽早识别并防止其发生。使用股骨加长指甲在八周内达到目标分散注意力的长度,并在进行性承重过程中保持机械衰竭。尝试重新分心,随后出现非负重状态。尽管患者无法达到最初的目标长度,但X射线照片显示患者的右下肢稳定缩短,并且骨骼再生不完全的早期迹象。她最终接受了用骨移植钉的交流钉。在重新评估髓内钉构造物的机械刚度后,应进一步研究和标准化术后方案和负重状态。