International intercomparison exercise of active radon devices and passive detectors at the First East European Radon Symposium (FERAS 2012)

机译:在第一次东欧Rad研讨会上进行的有源ra设备和无源探测器的国际比对活动(FERAS 2012)

获取原文
           

摘要

During the First East European Radon Symposium (2012), an Intercomparison Exercise of active devices and passive detectors was organized, at two exposure levels. Medium Radon Exposure was made in the laboratory room at low to medium level by an average radon concentration of hundreds of Bq·m-3 and involved 12 active devices and 8 groups of passive detectors. High Radon Exposure was performed in a calibration chamber at high level by an average radon concentration of thousands of Bq·m-3 and involved 5 active devices and 7 groups of passive detectors. Results of active devices show that mean concentrations of 11 devices from the Medium Exposure and 1 device from High Exposure were inside ± 95% C.I of the means. Results of passive detectors show that the exposure values of 4 groups from the Medium Exposure and 6 groups from the High Exposure were inside ± 20 % admittance level of the reference exposure.
机译:在首届东欧Rad研讨会(2012年)期间,组织了一次有源设备和无源探测器的比对练习,有两种暴露水平。中等room暴露是在实验室室内以低至中等水平进行的,平均ra浓度为数百Bq·m-3,涉及12个有源器件和8组无源探测器。高Rad暴露是在校准室内高水平进行的,平均m浓度为数千Bq·m-3,涉及5个有源器件和7组无源探测器。有源设备的结果表明,中等浓度的11种设备和高浓度的1种设备的平均浓度在平均值的95%C.I以内。无源检测器的结果表明,中度曝光的4组和高曝光度的6组的曝光值均在参考曝光的准入水平的±20%范围内。