首页> 外文期刊>South African medical journal = >Validation of the CoaguChek XS international normalised ratio point-of-care analyser in patients at Charlotte Maxeke Johannesburg Academic Hospital, South Africa
【24h】

Validation of the CoaguChek XS international normalised ratio point-of-care analyser in patients at Charlotte Maxeke Johannesburg Academic Hospital, South Africa

机译:南非夏洛特·马克斯·约翰内斯堡学术医院患者的CoaguChek XS国际标准化比率即时分析仪的验证

获取原文
           

摘要

BACKGROUND: Measurement of the international normalised ratio (INR) is essential in the management of patients on long-term warfarin therapy. The CoaguChek XS portable coagulometer is a point-of-care test for INR measurement. It offers the advantage of improved patient accessibility, particularly in peripheral clinics. OBJECTIVES: To evaluate the clinical utility of the CoaguChek XS for monitoring of patients on standard warfarin therapy (INR 2 - 3) as well as those with mechanical heart valve replacements (INR 2.5 - 3.5). METHODS:We compared the performance of the CoaguChek XS device with that of the STAGO laboratory analyser with regard to accuracy and precision in 304 patients referred for routine testing. RESULTS: The mean INR value of the CoaguChek XS of 2.75 (standard deviation (SD) 1.18) was comparable to that of the STAGO (2.65 (SD 1.04)). The Bland-Altman difference plot revealed good agreement. Bias between the two methods was small, and the imprecision was within acceptable limits. Within the target range (2.0 - 3.5), 93.9% of the CoaguChek XS INR readings were within 0.5 units of the standard laboratory method result. There was, however, an increase in the variability of the differences between the two test methods when the INR was 3.6. CONCLUSION: The CoaguChek XS point-of-care device can be used to provide accurate and precise INR measurements over a wide range for monitoring of valvular and non-valvular patients on long-term warfarin therapy.
机译:背景:国际标准化比率(INR)的测量对于长期接受华法林治疗的患者至关重要。 CoaguChek XS便携式血凝仪是用于INR测量的即时检验。它具有改善患者可及性的优势,特别是在外围诊所。目的:评估CoaguChek XS在监测标准华法林治疗(INR 2-3)和机械瓣膜置换术(INR 2.5-3.5)的患者中的临床效用。方法:我们比较了CoaguChek XS设备和STAGO实验室分析仪在304例接受常规检测的患者的准确性和精密度方面的性能。结果:CoaguChek XS的平均INR值为2.75(标准差(SD)1.18)与STAGO的平均INR值(2.65(SD 1.04))相当。 Bland-Altman差异图显示出良好的一致性。两种方法之间的偏差很小,并且不精确度在可接受的范围内。在目标范围内(2.0-3.5),CoaguChek XS INR读数的93.9%在标准实验室方法结果的0.5个单位内。但是,当INR> 3.6时,两种测试方法之间差异的变异性增加。结论:CoaguChek XS即时医疗设备可用于在广泛范围内提供准确,精确的INR测量,以监测长期接受华法林治疗的瓣膜和非瓣膜患者。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号