Linearity Test for Harshaw TLD (Type: TLD-100H) Base on Individual Calibration Method

机译:基于单独校准方法的Harshaw TLD(类型:TLD-100H)线性度测试

获取原文
       

摘要

The testing of the individual monitoring instruments is important to demonstrate the performance of the instruments to give accurate measurements in workplace environment. In this research, 18 Thermoluminescence dosimetry (TLD) units were calibrated individually at surface water phantom and exposed with ~(60)Co source at block 32 in Malaysia Nuclear Agency. The TLD were exposed at 5.00 meter distance from the source. The exposed TLD in terms of Personal Dose Equivalent at 10mm depth tissue, (Hp (10)) equal to 2.00mSv. The exposed TLD then be measured using winRems software from Harshaw TLD reader 6600 plus for defining the calibration factor in term of mSvC. After that all the 18 unit TLD were tested using linearity testing method and 18 TLD units were exposed with different dose that were 1mSv, 5mSv, 7mSv, 10mSv, 15mSv, and 20mSv. The research is conducted to satisfy two main objectives which was to obtain linear regression coefficient R~2 ~ 1 and to show that the ratio of measured value over standard values are within ICRP trumpet acceptance limit curve, which are within (-33% to +50%).
机译:单独监视仪器的测试对于证明仪器的性能非常重要,以便在工作场所环境中进行准确的测量。在这项研究中,在地表水体模上分别校准了18个热致发光剂量法(TLD)单元,并在马来西亚核机构的第32区块中使用〜(60)Co源进行了曝光。 TLD在离源5.00米的距离处暴露。就10mm深度组织的个人剂量当量而言,暴露的TLD(Hp(10))等于2.00mSv。然后使用Harshaw TLD阅读器6600 plus的winRems软件测量暴露的TLD,以定义以mSv / nC表示的校准因子。之后,使用线性测试方法测试所有18个TLD单元,并分别以1mSv,5mSv,7mSv,10mSv,15mSv和20mSv的不同剂量暴露18个TLD单元。进行研究以满足两个主要目标,即获得线性回归系数R〜2〜1,并表明测量值与标准值之比在ICRP小号接受极限曲线内,且在(-33%至+ 50%)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号