Comparison of Total Bilirubin Values Measured with ABL 735 Blood Gas Analyzer and Roche Cobas C8000 Chemistry Analyzer in Age-Segregated Pediatric Patients

机译:使用ABL 735血气分析仪和Roche Cobas C8000化学分析仪对年龄不同的小儿患者测得的总胆红素值的比较

获取原文
       

摘要

Aim: Measurement of blood bilirubin levels is a crucial analysis because of the toxic effects of bilirubin on brain tissue, particularly in preterm neonates. The aim of this study was to investigate the consistency of the total bilirubin values obtained by the blood gas analyzer and the autoanalyzer.Material and Methods: In this study, we used total bilirubin data of 407 pediatric patients from Kocaeli University Medical Faculty Education and Research Hospital Central Laboratory System. Total bilirubin data, provided that it was measured simultaneously, was obtained from ABL 735 blood gas analyzer and Roche Cobas C8000 chemistry analyzer. Pediatric patients (neonates, infant and children under 17 years old) were selected retrospectively by year between 2015-2017.Results: Under a cut-off value (14.6 mg/dL) ABL 735 blood gas analyzer and Roche COBAS C8000 chemistry analyzer had strong correlation (r = 0.939) for total bilirubin measurements. It was found that 2-15 days old neonates give more scattered total bilirubin data by Bland Altman analysis in two measurements. Statistical analysis performed to compare whole total bilirubin data identity between two measurements: correlation coefficient was found r = 0.949 a statistically significant positive correlation (p < 0.001).Conclusion: According to our analysis which was supported by previous studies in the literature, we can say that the compatibility between the blood gas analyzer (multi-wave-length spectrophotometric technique) and the chemistry analyzer becomes weaker when the total bilirubin levels exceed 14.6 mg/dL.
机译:目的:由于胆红素对脑组织,特别是早产儿的毒性作用,血胆红素水平的测量是至关重要的分析。这项研究的目的是研究通过血气分析仪和自动分析仪获得的总胆红素值的一致性。材料与方法:本研究中,我们使用了Kocaeli大学医学院教与学的407名儿科患者的总胆红素数据。医院中央实验室系统。如果同时进行测量,总胆红素数据可从ABL 735血气分析仪和Roche Cobas C8000化学分析仪获得。在2015年至2017年之间按年龄对儿童患者(新生儿,婴儿和17岁以下儿童)进行了回顾性研究。结果:在临界值(14.6 mg / dL)下,ABL 735血气分析仪和Roche COBAS C8000化学分析仪具有很强的抵抗力总胆红素测量值的相关性(r = 0.939)。通过Bland Altman分析在两次测量中发现,2-15天大的新生儿提供了更多分散的总胆红素数据。进行统计分析以比较两次测量之间的总胆红素数据的整体同一性:相关系数为r = 0.949,具有统计学显着正相关(p <0.001)。结论:根据我们的分析,该分析得到了先前文献的支持,我们可以有人说,当总胆红素水平超过14.6 mg / dL时,血气分析仪(多波长分光光度技术)与化学分析仪之间的兼容性变弱。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号