首页> 外文期刊>Radiation oncology >Dosimetric comparison using different multileaf collimeters in intensity-modulated radiotherapy for upper thoracic esophageal cancer
【24h】

Dosimetric comparison using different multileaf collimeters in intensity-modulated radiotherapy for upper thoracic esophageal cancer

机译:强度调制放射治疗上胸段食管癌中使用不同多叶测径仪进行剂量学比较

获取原文
           

摘要

Purpose To study the impacts of multileaf collimators (MLC) width [standard MLC width of 10 mm (sMLC) and micro-MLC width of 4 mm (mMLC)] in the intensity-modulated radiotherapy (IMRT) planning for the upper thoracic esophageal cancer (UTEC). Methods and materials 10 patients with UTEC were retrospectively planned with the sMLC and the mMLC. The monitor unites (MUs) and dose volume histogram-based parameters [conformity index (CI) and homogeneous index (HI)] were compared between the IMRT plans with sMLC and with mMLC. Results The IMRT plans with the mMLC were more efficient (average MUs: 703.1 ± 68.3) than plans with the sMLC (average MUs: 833.4 ± 73.8) (p p 5 (3260.3 ± 374.0 vs 3404.5 ± 374.4)/gEUD (1815.1 ± 281.7 vs 1849.2 ± 297.6) of the spinal cord, the V10 (33.2 ± 6.5 vs 34.0 ± 6.7), V20 (16.0 ± 4.6 vs 16.6 ± 4.7), MLD (866.2 ± 174.1 vs 887.9 ± 172.1) and gEUD (938.6 ± 175.2 vs 956.8 ± 171.0) of the lungs were observed in the plans with the mMLC, respectively (p Conclusions Comparing to the sMLC, the mMLC not only demonstrated higher efficiencies and more optimal target coverage, but also considerably improved the dose sparing of OARs in the IMRT planning for UTEC.
机译:目的研究多叶准直仪(MLC)宽度[标准MLC宽度10 mm(sMLC)和micro-MLC宽度4 mm(mMLC)]对上胸段食管癌的强度调节放疗(IMRT)规划的影响(UTEC)。方法和材料回顾性规划sMLC和mMLC对10例UTEC患者。在带有sMLC和带有mMLC的IMRT计划之间,比较了监测单位(MUs)和基于剂量体积直方图的参数[合格指数(CI)和均质指数(HI)]。结果使用mMLC的IMRT计划(平均MU:703.1±68.3)比使用sMLC的IMRT计划(平均MU:833.4±73.8)(pp 5 (3260.3±374.0 vs 3404.5±374.4)/ gEUD(脊髓的1815.1±281.7与1849.2±297.6),V 10 (33.2±6.5与34.0±6.7),V 20 (16.0±4.6与16.6±4.7) ),在使用mMLC的计划中分别观察到MLD(866.2±174.1 vs 887.9±172.1)和gEUD(938.6±175.2 vs 956.8±171.0)(p结论与sMLC相比,mMLC不仅表现出更高的肺功能)效率和更佳的目标覆盖率,但也大大改善了UTEC的IMRT规划中OAR的剂量节约。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号