Ambient radiation levels in positron emission tomography/computed tomography (PET/CT) imaging center

机译:正电子发射断层扫描/计算机断层扫描(PET / CT)成像中心的环境辐射水平

获取原文
       

摘要

Objective: To evaluate the level of ambient radiation in a PET/CT center. Materials and Methods: Previously selected and calibrated TLD-100H thermoluminescent dosimeters were utilized to measure room radiation levels. During 32 days, the detectors were placed in several strategically selected points inside the PET/CT center and in adjacent buildings. After the exposure period the dosimeters were collected and processed to determine the radiation level. Results: In none of the points selected for measurements the values exceeded the radiation dose threshold for controlled area (5 mSv/year) or free area (0.5 mSv/year) as recommended by the Brazilian regulations. Conclusion: In the present study the authors demonstrated that the whole shielding system is appropriate and, consequently, the workers are exposed to doses below the threshold established by Brazilian standards, provided the radiation protection standards are followed.
机译:目的:评估PET / CT中心的环境辐射水平。材料和方法:使用先前选择和校准的TLD-100H热辐射剂量计来测量房间的辐射水平。在32天中,将探测器放置在PET / CT中心内部和相邻建筑物中的几个关键位置。在暴露期之后,收集剂量计并进行处理以确定辐射水平。结果:在所有未选择的测量点中,其值均未超出巴西法规建议的控制区域(5 mSv /年)或自由区域(0.5 mSv /年)的辐射剂量阈值。结论:在本研究中,作者证明了整个屏蔽系统是适当的,因此,只要遵循辐射防护标准,工人的暴露剂量应低于巴西标准规定的阈值。