An assessment of scattered radiation during fluoroscopic procedures in diagnostic radiology

机译:诊断放射学中荧光检查过程中的散射辐射评估

获取原文
       

摘要

The results of measurements of scattered radiation in the vicinity of a fluoroscopic X-ray facility are presented in this paper. Two different fluoroscopic systems, one with an undercouch tube and one with an overcouch tube, were compared. The dose rate was measured during the simulation of a fluoroscopy procedure, using an ionization chamber as a dosemeter. The distribution of scattered radiation has been determined and results show a much higher dose rate in cases of an overcouch tube arrangement. When X-ray units with an undercouch tube are concerned, under same exposure conditions, the dose rate is higher in cases of a vertical beam. Prior to the measurements, the ionization chamber was examined in order to evaluate its suitability as a survey meter used in diagnostic radiology. Measurements show that below 1.2 s, the ionization chamber gives an underestimation of dose rates. Therefore, in order to perform accurate measurements using this instrument, exposure times should be above 1.2 s.
机译:本文介绍了X射线透视设备附近散射辐射的测量结果。比较了两种不同的荧光镜系统,其中一种带有下沙发管,另一种带有上沙发管。在荧光检查程序的模拟过程中,使用电离室作为剂量计,测量剂量率。已经确定了散射辐射的分布,并且在长沙发管布置的情况下,结果显示出更高的剂量率。当考虑到带有下沙发管的X射线装置时,在相同的曝光条件下,垂直光束的剂量率更高。在测量之前,对电离室进行了检查,以评估其作为诊断放射学中使用的测量仪的适用性。测量表明,在1.2 s以下,电离室会低估剂量率。因此,为了使用本仪器执行准确的测量,曝光时间应在1.2 s以上。