Risperdal ? CONSTA ? Needle Detachment. Incidence Rates Before and After Kit Redesign: A Retrospective Study using Electronic Health Records and Natural Language Processing in the Department of Veterans Affairs

机译:利培达? CONSTA?针头分离。重新设计套件之前和之后的发病率:使用退伍军人事务部的电子病历和自然语言处理进行的回顾性研究

获取原文
   

获取外文期刊封面封底 >>

       

摘要

IntroductionJanssen received reports of needle detachments for Risperdalsup?/sup CONSTAsup?/sup and, in response, redesigned the kit. ObjectiveThe study objective was to estimate the rate of Risperdalsup?/sup CONSTAsup?/sup needle detachments prior to and after the introduction of a redesigned kit. MethodsThis retrospective study used record abstraction in the US Department of Veterans Affairs (VA). The 3 phases included: (1) a pilot study for methods evaluation in a sample of 6 hospitals with previously reported detachments; (2) a baseline study to ascertain the baseline detachment rate; and (3) a follow-up study to ascertain the rate for the redesigned kit. Administrative codes and natural language processing with clinical review were used to identify detachments. ResultsPilot: we identified a subset of spontaneously reported detachments and several previously unreported events. In the baseline study (original device), from January through December 2013, 22 needle detachments were identified among 47,934 administrations of the drug in a census of administrations in the VA; an incidence of 0.0459%. In the follow-up study (redesigned device), from December 2015 through December 2016, there were 14 reported detachments in 41,819 injections, 0.0335%. This represents a reduction of 27% from the baseline. ConclusionThis approach enabled us to identify needle detachments we would not have otherwise found (“solicited”). However, it likely resulted in incomplete outcome ascertainment. While this may have resulted in lower overall rates, it did not bias the comparison of the baseline and follow-up studies. The results showed that the redesigned Risperdalsup?/sup CONSTAsup?/sup kit reduced the incidence of needle detachment events in the VA. FundingJanssen Pharmaceuticals, Inc.
机译:简介Janssen收到了有关Risperdal ? CONSTA ?的针头脱落的报告,并为此进行了重新设计。目的研究目的是评估在引入重新设计的试剂盒之前和之后Risperdal ? CONSTA ?针头脱离的发生率。方法:这项回顾性研究使用了美国退伍军人事务部(VA)的记录摘要。这三个阶段包括:(1)对6家医院的样本进行过方法评估的前瞻性研究,这些医院以前曾报告过分遣队; (2)基线研究以确定基线脱离率; (3)进行后续研究,以确定重新设计套件的价格。使用行政法规和自然语言处理并进行临床审查来识别支队。结果飞行员:我们确定了自发报告的超脱子集和一些以前未报告的事件。在基线研究(原始设备)中,从2013年1月至2013年12月,在弗吉尼亚州的一次普查中,在47,934剂药物中发现了22个针头脱离;发生率为0.0459%。在2015年12月至2016年12月的后续研究(重新设计的装置)中,报告的14 819次注射中有14例脱离,占0.0335%。与基线相比减少了27%。结论这种方法使我们能够识别原本不会发现的针头分离(“主动”)。但是,这可能导致结果确认不完整。尽管这可能会降低总体发病率,但它不会使基线研究和随访研究的比较产生偏差。结果表明,重新设计的Risperdal ? CONSTA ?试剂盒减少了VA中针头脱落事件的发生。资金詹森制药有限公司

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号