Comparison of Goldmann Applanation, Noncontact Air Puff, and Tono-Pen XL Tonometry in Normal Controls versus Glaucoma Patients at a University Hospital in Riyadh, Saudi Arabia

机译:沙特阿拉伯利雅得大学医院正常对照与青光眼患者的戈德曼压平法,非接触式吹气法和Tono-Pen XL眼压计的比较

获取原文
           

摘要

PURPOSE: The purpose of this study is to compare intraocular pressure (IOP) measured by the Goldmann applanation tonometer (GAT), the Tono-Pen XL (TPXL), and a noncontact airpuff tonometer (NCT) in glaucoma patients and normal controls. METHODOLOGY: In the current study, two groups of individuals were recruited; the first group included glaucoma patients (glaucoma group) while the second one was a glaucoma-free group. The IOP was measured through GAT, NCT, and TPXL for the same participants by three different physicians in both groups. Measurements through the three devices were compared statistically. Correlations between different methods of assessment were also assessed. RESULTS: In the glaucoma group, the mean IOP measured was 16.0 ± 5.5 mmHg (range, 6–40 mmHg) with GAT 20.5 ± 6.9 mmHg (range, 10–52 mmHg) with the TPXL, and 20.2 ± 6.5 mmHg (range, 8–50 mmHg) with the NCT. In the control group, the mean IOP was 14.0 ± 2.7 mmHg (range, 9–19 mmHg) with GAT, 17.3 ± 3.8 mmHg (range 6–30 mmHg) with the TPXL, and 17.9 ± 3.9 mHg (range 10–27 mmHg) with the NCT. CONCLUSION: IOP measurements were approved among the three devices with relatively higher readings using both NTC and TPXL. All three methods are required to address different situations that present in the daily clinical and surgical practice.
机译:目的:本研究的目的是比较青光眼患者和正常对照者的戈德曼压平眼压计(GAT),Tono-Pen XL(TPXL)和非接触气喘眼压计(NCT)测量的眼内压(IOP)。方法:在本研究中,招募了两组个人;分别为第一组包括青光眼患者(青光眼组),而第二组是无青光眼组。两组中三位不同的医师通过GAT,NCT和TPXL对同一参与者的IOP进行了测量。通过统计比较了通过这三个设备的测量结果。还评估了不同评估方法之间的相关性。结果:青光眼组的平均IOP为16.0±5.5 mmHg(范围6–40 mmHg),GAT为20.5±6.9 mmHg(范围10–52 mmHg),而TPXL为20.2±6.5 mmHg(范围, 8–50 mmHg)。在对照组中,GAT的平均IOP为14.0±2.7 mmHg(范围9–19 mmHg),TPXL的平均IOP为17.3±3.8 mmHg(范围6–30 mmHg),TPS的平均IOP为17.9±3.9 mHg(范围10–27 mmHg) )与NCT。结论:在NTC和TPXL两种读数相对较高的设备中,均批准了IOP测量。需要使用这三种方法来解决日常临床和手术实践中出现的不同情况。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号