首页> 外文期刊>Microbiologia Medica >Effect evaluation on a medical device for drinking water filtration designed for reduction of microbial density in areas at risk
【24h】

Effect evaluation on a medical device for drinking water filtration designed for reduction of microbial density in areas at risk

机译:对用于降低危险区域中微生物密度的饮用水过滤医疗设备的效果评估

获取原文
   

获取外文期刊封面封底 >>

       

摘要

In areas at risk (hospitals, nursing homes) but in also in community such as schools, swimming pools, gyms, drinking water plumbing system can harbor also pathogenic microrganisms, potential vehicles of diseases because of structural factors and phenomena of microbial regrowth. In a setting of prevention, for limiting the exposure to pathogens in the environment and for minimizing the risk, at the point of use, specific disposable filter units could be installed. Aims. The purpose of this study was to examine the effectiveness in retaining microrganisms of a medical devices installed at drinking water faucets. Materials and methods. One hundred and eighty samples of water were microbiologically analyzed. Natural heterotrophic microbial flora present in drinking water were analytically determined before and after the passage through the system. Contemporaneously, twenty four suspensions of Pseudomonas aeruginosa and Escherichia coli at known concentration were determined for the calculation of the limit of retention calculated taking into examination the worst cases associated with the highest microbial concentrations. Results. The investigation showed that the devices were both efficient at reducing the natural microbial load and at guaranteeing the removal of high concentration of microrganisms (till 5 log). Fifty-seven out of 90 samples analyzed for the heterotrophs did not show any growth after the drainage through the system. In the positive samples, 86% showed a microbial concentration ranging between 1 e 3 CFU/L. Progressive growing concentrations of P. aeruginosa and E. coli proved a high limit of retention: none growth was obtained till the concentrations of 1.3x1012 CFU/mL and 5.0x1013 CFU/mL, respectively. Conclusions. The data show that the filter units installed directly at the point of use can be helpful at guaranteeing reliable retention of microbes. Still, in order to obtain an appropriate operation of the device, in spite of both the ease of use and safety margin, it is necessary to implement an efficient and orderly activity of maintenance according to the manufacturer’s instructions.
机译:在高危地区(医院,疗养院)但在社区中(如学校,游泳池,体育馆),饮用水管道系统也可能带有致病微生物,由于结构因素和微生物再生现象,可能成为疾病的传播媒介。在预防方面,为了限制环境中病原体的暴露并最大程度地降低风险,可以在使用时安装特定的一次性过滤器单元。目的这项研究的目的是检查保留在饮用水水龙头上的医疗设备中微生物的有效性。材料和方法。对180份水进行了微生物学分析。在通过系统之前和之后,对饮用水中存在的天然异养微生物菌群进行了分析测定。同时,确定24种已知浓度的铜绿假单胞菌和大肠杆菌的悬浮液,以计算保留限,并考虑了与最高微生物浓度相关的最坏情况。结果。调查表明,该设备既可以有效降低自然微生物负荷,又可以保证去除高浓度的微生物(至5 log)。在分析了异养菌的90个样品中,有57个没有显示出通过系统排水后的任何增长。在阳性样本中,有86%的微生物浓度范围为1 e 3 CFU / L。逐渐增加的铜绿假单胞菌和大肠杆菌的浓度被证明具有很高的保留限度:直到分别达到1.3x1012 CFU / mL和5.0x1013 CFU / mL的浓度,都没有生长。结论。数据表明,直接在使用时安装的过滤器有助于确保可靠地保留微生物。尽管如此,为了获得设备的适当操作,尽管易于使用且具有安全裕度,但仍必须根据制造商的说明进行有效而有序的维护活动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号