首页> 外文期刊>MethodsX >Updating the ELISA standard curve fitting process to reduce uncertainty in estimated microcystin concentrations
【24h】

Updating the ELISA standard curve fitting process to reduce uncertainty in estimated microcystin concentrations

机译:更新ELISA标准曲线拟合过程以减少估计的微囊藻毒素浓度的不确定性

获取原文

摘要

This study is aimed at exploring the optimal ELISA standard curve fitting process for reducing measurement uncertainty. Using an ELISA kit for measuring cyanobacterial toxin (microcystin), we show that uncertainty associated with the estimated microcystin concentrations can be reduced by defining the standard curve as a four-parameter logistic function on the natural log concentration scale, instead of the current approach of defining the curve on the concentration scale. The model comparison method is outlined in this paper, allowing it to be transferable to test different statistical models for other ELISA test kits.
机译:这项研究旨在探索减少测量不确定度的最佳ELISA标准曲线拟合过程。使用ELISA试剂盒测量蓝藻毒素(微囊藻毒素),我们显示出可以通过将标准曲线定义为自然对数浓度标度上的四参数对数函数来减少与估计的微囊藻毒素浓度相关的不确定性,而不是使用当前方法在浓度标度上定义曲线。本文概述了模型比较方法,该方法可用于测试其他ELISA测试试剂盒的不同统计模型。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号