Background: In radiotherapy treatments, it is crucial to monitor the performance of linac components including gantry, collimation system, and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments. Methods: The EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in leaf bank assembly due to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with five ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of image data. Three Siemens linacs were investigated. Results: The average EPID sag was within 1 mm for all tested linacs. Two machines showed >1 mm gantry sag. Changes in the SDD values were within 7.5 mm. EPID skewness and tilt values were <1° in all machines. The maximum sag in leaf bank assembly was <1 mm. Conclusion: The method and software developed in this study provide a simple tool for effective investigation of the behavior of Siemens linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Siemens machines.
机译:背景:在放射治疗中,至关重要的是在电弧输送过程中监控直线加速器组件的性能,包括机架,准直系统和电子门成像设备(EPID)。在这项研究中,提出了一种基于EPID的简单测量方法,并结合一种算法来研究这些系统在不同机架角度的稳定性,目的是评估与治疗机相关的误差。方法:通过采集37张EPID图像,分别研究了EPID凹陷,龙门凹陷,源到检测器距离(SDD)的变化,EPID和准直仪偏斜度,EPID倾斜以及直线加速器旋转导致的叶岸组件中的凹陷。一个简单的幻影,带有五个在不同机架角度的球轴承。开发了一种快速而强大的软件包来自动分析图像数据。研究了三种西门子直线加速器。结果:所有测试的直线加速器的平均EPID下陷均在1毫米内。两台机器显示了> 1毫米的龙门下陷。 SDD值的变化在7.5毫米内。在所有机器中,EPID偏度和倾斜值均<1°。叶库组件中的最大下垂<1 mm。结论:本研究开发的方法和软件为有效研究带有门架旋转的西门子直线加速器组件行为提供了简单的工具。如此全面的研究首次在西门子机器上进行。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号