首页> 外文期刊>Medical Devices: Evidence and Research >New bipolar tissue ligator combines constant tissue compression and temperature guidance: histologic study and implications for treatment of hemorrhoids
【24h】

New bipolar tissue ligator combines constant tissue compression and temperature guidance: histologic study and implications for treatment of hemorrhoids

机译:新型双极组织结扎器结合了恒定的组织压缩和温度指导:组织学研究及其对痔疮的治疗意义

获取原文
           

摘要

Background: Several minimally invasive technologies are available to treat common soft tissue lesions including symptomatic hemorrhoids. The use of energy to deliver heat and coagulate target lesions is commonly practiced. This study compares the histologic effects produced on intestinal tissues by two energy-based systems which employ different approaches of heat delivery.Methods: Two heat delivery systems were evaluated in vivo in a single porcine subject: infrared coagulator and bipolar tissue ligator utilizing constant tissue compression and temperature guidance. Eighteen treatment sites divided into three groups of six were assessed. Treatment site temperature was measured and the effects of thermal treatment in the mucosa, submucosa, submucosal vessels, and muscularis layer were scored. Lateral thermal spread beyond the energy application site was also assessed.Results: Treatment site temperatures were much lower in the bipolar ligator group than in the infrared coagulator group. The mucosal and submucosal tissue changes observed in tissues treated with infrared energy and bipolar energy at 55°C were similar. Both the mucosal and submucosal tissue changes with bipolar energy at 50°C were significantly less.Conclusion: Both devices achieved similar histologic results. However, the unique design of the bipolar ligator, which allows consistent capture, constant compression, and temperature monitoring of target tissue, accomplished the desired histologic changes with less muscular damage at much lower temperatures than the infrared coagulator. The use of bipolar ligation could offer clinical advantages such as reduced patient pain and a minimized chance of heat-related collateral tissue damage.
机译:背景:几种微创技术可用于治疗常见的软组织病变,包括有症状的痔疮。通常实践使用能量来传递热量和凝结靶病变。本研究比较了两种采用不同热传递方式的能量系统对肠道组织产生的组织学影响。方法:在单个猪受试者体内对两种热传递系统进行了评估:红外凝结器和利用恒定组织压缩的双极组织结扎器和温度指导。评估了18个治疗部位,分为3组,每组6个。测量治疗部位的温度,并对粘膜,粘膜下层,粘膜下血管和肌层的热处理效果进行评分。结果:双极结扎器组的治疗部位温度比红外凝结器组低得多。在55℃下用红外能量和双极能量处理的组织中观察到的粘膜和粘膜下组织变化是相似的。在50°C时,双极能量对粘膜和粘膜下组织的改变均明显减少。结论:两种装置均获得了相似的组织学结果。但是,双极结扎器的独特设计可实现对目标组织的持续捕获,恒定压缩和温度监控,从而在比红外凝结器低得多的温度下以较少的肌肉损伤实现了所需的组织学改变。双极结扎术的使用可提供临床优势,例如减轻患者的痛苦,并减少与热相关的副组织损伤的机会。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号