首页> 外文期刊>Frontiers in Veterinary Science >Do the Manual or Computer-Controlled Flowmeters Generate Similar Isoflurane Concentrations in Tafonius?
【24h】

Do the Manual or Computer-Controlled Flowmeters Generate Similar Isoflurane Concentrations in Tafonius?

机译:手动流量计或计算机控制的流量计在Tafonius中会产生相似的异氟醚浓度吗?

获取原文

摘要

Introduction: Tafonius is an anaesthesia machine with computer-controlled monitor and ventilator. We compared the isoflurane fluctuations in the circuit with manual (MF) or computer-driven (CF) flowmeters, investigated the origin of the differences and assessed whether isoflurane concentration time course followed a one-compartment model. Material and Methods: A calibrated TEC-3 isoflurane vaporiser was used. Gas composition and flows were measured using a multiparametric monitor and a digital flowmeter. Measurements included: 1) Effects of various FiO2 with MF/CF on the isoflurane fraction changes in the breathing system during mechanical ventilation of a lung model; wash-in kinetic was fitted to a compartmental model; 2) Gas outflow at the common gas outlet (CGO) with MF/CF at different FiO2; 3) Isoflurane output of the vaporiser at various dial settings with MF/CF set at different flows without and with reduction of the CGO diameter. Results: 1) The 3% targeted isoflurane concentration was not reached; additional time was required to reach specific concentrations with CF (lowest FiO2, longer time). The exponential course fitted a two-compartment model; 2) Set and measured flows were identical with MF. With CF at 0.21 FiO2, flow was intermittently 7.6 L minute-1 or zero (mean total: 38% of the set flow); with CF at 1.00 FiO2, flow was 10.6 L minute-1 or zero (mean: 4-5.3 L minute-1); with 0.21FiO21.00, combined flow was intermittent (maximum output: 15.6 L minute-1); 3) With MF, isoflurane output was matching dial setting at 5 L minute-1 but was lower at higher flows; with CF generating intermittent flows, isoflurane output was fluctuating. With the 4 mm diameter CGO, isoflurane concentration was close to dial setting with both MF and CF. With a 14 G CGO, isoflurane concentration was lower than dial setting with MF, higher with CF. Conclusions and clinical relevance: Using MF or CF led to different isoflurane fraction time course in Tafonius. Flows were lower than set with CF; the TEC-3 did not compensate for high/intermittent flows and pressures; the CGO diameter influenced isoflurane output.
机译:简介:Tafonius是一种麻醉机,带有计算机控制的监视器和呼吸机。我们使用手动(MF)或计算机驱动(CF)流量计比较了回路中的异氟醚波动,调查了差异的根源,并评估了异氟醚浓度时程是否遵循一室模型。材料和方法:使用校准的TEC-3异氟烷蒸发器。使用多参数监视器和数字流量计测量气体成分和流量。测量包括:1)在肺部模型的机械通气期间,各种MF / CF混合的FiO2对呼吸系统中异氟烷组分变化的影响;冲洗动力学适合于隔室模型; 2)在公共气体出口(CGO)处的气体流出,其中MF / CF在不同的FiO2中; 3)在不同的刻度盘设置下,气化器的异氟烷输出,MF / CF设置为不同的流量,且不降低CGO直径。结果:1)异氟醚目标浓度未达到3%;用CF达到特定浓度需要额外的时间(最低的FiO2,更长的时间)。指数过程符合两室模式。 2)设定流量和测量流量与MF相同。 CF为0.21 FiO2时,流量间歇地为7.6 L min-1或为零(平均:设定流量的38%); CF为1.00 FiO2时,流量为10.6 L分钟-1或为零(平均值:4-5.3 L分钟-1);在0.21

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号